Rhoden D L, O'Hara C M
Nosocomial Infections Laboratory Branch, Centers for Disease Control, Atlanta, Georgia 30333.
J Clin Microbiol. 1989 Nov;27(11):2420-2. doi: 10.1128/jcm.27.11.2420-2422.1989.
The QUANTUM II system (Abbott Diagnostics, Irving, Tex.) was evaluated with 65 species of gram-negative bacilli from various culture collections at the Centers for Disease Control. The QUANTUM II system accurately identified 92.5% of 335 isolates tested, as follows: 92.6% of 258 members of the family Enterobacteriaceae, 92.7% of 55 nonfermenters, and 91% of 22 oxidase-positive fermenters. These results were obtained by using the additional biochemical and serologic tests recommended by the manufacturers of the QUANTUM II system. The 25 misidentified cultures generally belonged to newly recognized genera, atypical strains, or slower-growing strains of more widely known genera. The system identified the most commonly encountered organisms at an accuracy of greater than or equal to 95%. The system is efficient, accurate, and rapid.
对量子II系统(雅培诊断公司,得克萨斯州欧文市)进行了评估,使用了美国疾病控制中心各种培养物保藏中心的65种革兰氏阴性杆菌。量子II系统准确鉴定了所测试的335株分离菌中的92.5%,具体如下:肠杆菌科258个成员中的92.6%,非发酵菌55株中的92.7%,以及氧化酶阳性发酵菌22株中的91%。这些结果是通过使用量子II系统制造商推荐的额外生化和血清学检测获得的。25株误鉴定的培养物通常属于新确认的属、非典型菌株或更广为人知的属中生长较慢的菌株。该系统以大于或等于95%的准确率鉴定了最常见的微生物。该系统高效、准确且快速。