Seeberg Trine M, Orr James G, Opsahl Hanne, Austad H O, Roed Morten H, Dalgard Steffen H, Houghton David, Jones David E J, Strisland Frode
IEEE Trans Biomed Eng. 2017 Jul;64(7):1469-1478. doi: 10.1109/TBME.2016.2606538. Epub 2016 Sep 12.
One promising approach for a continuous, noninvasive, cuff-less ambulatory blood pressure (BP) monitor is to measure the pulse wave velocity or the inversely proportional pulse transit time (PTT), based on electrical and optical physiological measurements in the chest area. A device termed IsenseU-BP+ has been developed for measuring continuous BP, as well as other physiological data. The objective of this paper is to present results from the first clinical evaluation with a wide range of patients. The study was set up to verify whether IsenseU-BP+ can be used to measure raw signals with sufficient quality to derive PTT. Methods: The test protocol, run 23 times on 18 different patients with nonalcoholic fatty liver disease, includes both supine measurement at rest as well as measurements during indoor cycling. Changes in PTT were compared with the BP changes measured using validated reference sensors. Results: IsenseU-BP+ measured signals with good quality during rest on 17 of 18 patients despite the high diversity in age, body shape, and body mass index. Evaluation during cycling was difficult due to a lack of good reference measurements.
IsenseU-BP+ measures PTT with high quality during supine rest and exercise and could therefore be suitable for deriving noninvasive continuous BP, although evaluation during exercise was limited due to inaccurate reference BP measurements.
Continuous, noninvasive measurement of BP would be highly beneficial in a number of clinical settings. Systems currently considered as the gold standard for the investigation of hypertension carry considerable limitations, which could be overcome by the method proposed here.
一种用于连续、无创、无袖带动态血压监测的有前景的方法是基于胸部区域的电学和光学生理测量来测量脉搏波速度或成反比的脉搏传输时间(PTT)。已开发出一种名为IsenseU-BP+的设备用于测量连续血压以及其他生理数据。本文的目的是展示对广泛患者进行首次临床评估的结果。该研究旨在验证IsenseU-BP+是否可用于测量质量足够高的原始信号以得出PTT。方法:该测试方案在18名患有非酒精性脂肪肝病的不同患者身上进行了23次,包括静息仰卧位测量以及室内骑行期间的测量。将PTT的变化与使用经过验证的参考传感器测量的血压变化进行比较。结果:尽管患者在年龄、体型和体重指数方面存在很大差异,但IsenseU-BP+在18名患者中的17名患者静息时测量的信号质量良好。由于缺乏良好的参考测量,骑行期间的评估很困难。
IsenseU-BP+在仰卧位休息和运动期间高质量地测量PTT,因此可能适用于得出无创连续血压,尽管由于参考血压测量不准确,运动期间的评估受到限制。
连续、无创的血压测量在许多临床环境中将非常有益。目前被视为高血压研究金标准的系统存在相当大的局限性,而此处提出的方法可以克服这些局限性。