Shrestha Prarthana, Kneepkens R, Vrijnsen J, Vossen D, Abels E, Hulsken B
Philips Digital Pathology Solutions, Veenpluis 6, Best, 5684 PC, The Netherlands.
J Pathol Inform. 2016 Dec 30;7:56. doi: 10.4103/2153-3539.197205. eCollection 2016.
The quality of images produced by whole slide imaging (WSI) scanners has a direct influence on the readers' performance and reliability of the clinical diagnosis. Therefore, WSI scanners should produce not only high quality but also consistent quality images.
We aim to evaluate reproducibility of WSI scanners based on the quality of images produced over time and among multiple scanners. The evaluation is independent of content or context of test specimen.
The ultimate judge of image quality is a pathologist, however, subjective evaluations are heavily influenced by the complexity of a case and subtle variations introduced by a scanner can be easily overlooked. Therefore, we employed a quantitative image quality assessment method based on clinically relevant parameters, such as sharpness and brightness, acquired in a survey of pathologists. The acceptable level of quality per parameter was determined in a subjective study. The evaluation of scanner reproducibility was conducted with Philips Ultra-Fast Scanners. A set of 36 HercepTest™ slides were used in three sub-studies addressing variations due to systems and time, producing 8640 test images for evaluation.
The results showed that the majority of images in all the sub-studies are within the acceptable quality level; however, some scanners produce higher quality images more often than others. The results are independent of case types, and they match our perception of quality.
The quantitative image quality assessment method was successfully applied in the HercepTest™ slides to evaluate WSI scanner reproducibility. The proposed method is generic and applicable to any other types of slide stains and scanners.
全玻片成像(WSI)扫描仪所产生图像的质量对读者的表现及临床诊断的可靠性有着直接影响。因此,WSI扫描仪不仅应产生高质量的图像,还应产生质量一致的图像。
我们旨在基于随时间推移以及多台扫描仪所产生图像的质量来评估WSI扫描仪的可重复性。该评估与测试标本的内容或背景无关。
图像质量的最终评判者是病理学家,然而,主观评估会受到病例复杂性的严重影响,扫描仪引入的细微变化很容易被忽视。因此,我们采用了一种基于临床相关参数(如清晰度和亮度)的定量图像质量评估方法,这些参数是在对病理学家的一项调查中获取的。每个参数的可接受质量水平是在一项主观研究中确定的。使用飞利浦超快速扫描仪对扫描仪的可重复性进行评估。在三项子研究中使用了一组36张HercepTest™玻片,以研究因系统和时间导致的变化,共产生8640张测试图像用于评估。
结果表明,所有子研究中的大多数图像都在可接受的质量水平内;然而,一些扫描仪比其他扫描仪更频繁地产生更高质量的图像。结果与病例类型无关,且与我们对质量的认知相符。
定量图像质量评估方法成功应用于HercepTest™玻片中,以评估WSI扫描仪的可重复性。所提出的方法具有通用性,适用于任何其他类型的玻片染色和扫描仪。