使用100份马血样本对一种新型手持式即时检验血气分析仪进行评估。

Evaluation of a new handheld point-of-care blood gas analyser using 100 equine blood samples.

作者信息

Bardell David, West Eleanor, Mark Senior J

机构信息

School of Veterinary Science, Leahurst Campus, University of Liverpool, Neston, UK.

School of Veterinary Science, Leahurst Campus, University of Liverpool, Neston, UK.

出版信息

Vet Anaesth Analg. 2017 Jan;44(1):77-85. doi: 10.1111/vaa.12392. Epub 2017 Feb 22.

Abstract

OBJECTIVES

To determine whether the Enterprise point-of-care blood analysis system (EPOC) produces results in agreement with two other blood gas analysers in regular clinical use (i-STAT and Radiometer ABL77) and to investigate the precision of the new machine when used with equine whole blood.

STUDY DESIGN

Prospective, randomized, non-blinded, comparative laboratory analyser study.

ANIMALS

Horses admitted to a university teaching hospital requiring arterial or venous blood gas analysis as part of their routine clinical management.

METHODS

One hundred equine blood samples were run immediately, consecutively and in randomized order on three blood gas analysers. Results of variables common to all three analysers were tested for agreement and compared with guidelines used in human medicine. These require 80% of results from the test analyser to fall within a defined range or percentage of results from the comparator devices to achieve acceptability. Additionally, 21 samples were run twice in quick succession on the EPOC analyser to investigate precision.

RESULTS

Agreement targets were not met for haematocrit, haemoglobin and base excess for either i-STAT or ABL77 analysers. EPOC precision targets were not met for partial pressure of carbon dioxide, ionized calcium, haematocrit and haemoglobin. Overall comparative performance of the EPOC was good to excellent for pH, oxygen tension, potassium, bicarbonate and oxygen saturation of haemoglobin, but marginal to poor for other parameters.

CONCLUSIONS AND CLINICAL RELEVANCE

The EPOC may be useful in performing analysis of equine whole blood, but trend analysis of carbon dioxide tension, ionized calcium, haematocrit and haemoglobin should be interpreted with caution. The EPOC should not be used interchangeably with other blood gas analysers.

摘要

目的

确定企业即时护理血液分析系统(EPOC)的检测结果是否与另外两种常规临床使用的血气分析仪(i-STAT和雷度ABL77)一致,并研究该新仪器用于马全血检测时的精密度。

研究设计

前瞻性、随机、非盲、比较实验室分析仪研究。

动物

入住大学教学医院、需要进行动脉或静脉血气分析作为常规临床管理一部分的马匹。

方法

100份马血样本在三种血气分析仪上立即、连续且随机排序进行检测。对所有三种分析仪共有的变量结果进行一致性测试,并与人类医学中使用的指南进行比较。这些指南要求测试分析仪80%的结果落在定义范围内,或落在比较设备结果的一定百分比范围内,才能达到可接受性。此外,21份样本在EPOC分析仪上快速连续检测两次,以研究精密度。

结果

i-STAT或ABL77分析仪的血细胞比容、血红蛋白和碱剩余均未达到一致性目标。EPOC分析仪在二氧化碳分压、离子钙、血细胞比容和血红蛋白方面未达到精密度目标。EPOC在pH值、氧分压、钾、碳酸氢盐和血红蛋白氧饱和度方面的总体比较性能良好至优秀,但在其他参数方面则为边缘至较差。

结论及临床意义

EPOC可能有助于进行马全血分析,但对二氧化碳分压、离子钙、血细胞比容和血红蛋白的趋势分析应谨慎解释。EPOC不应与其他血气分析仪互换使用。