Sandersen Charlotte, Dmitrovic Petra, Dupont Julien, Cesarini Carla, Guyot Hugues, Serteyn Didier, Kirsch Katharina
Department of Clinical Sciences, Faculty of Veterinary Medicine, University of Liege, 4000 Liege, Belgium.
Department of Clinical Sciences, Faculty of Veterinary Medicine, University of Zagreb, 10000 Zagreb, Croatia.
Vet Sci. 2023 Feb 3;10(2):114. doi: 10.3390/vetsci10020114.
Different blood gas analyzers are used in equine practice. Every machine needs to be validated, as they have not been designed for use in horses. The aim of this study was to compare the newly marketed GEM5000 machine to the formerly validated epoc machine for blood gas analysis in horses. In this prospective, non-blinded, comparative laboratory analyzer study, 43 equine blood samples were analyzed on both analyzers and values were compared between the two machines via Lin's concordance analysis, Passing-Bablok regression analysis and Bland-Altman plots. Duplicate measurements were conducted on the GEM5000 machine to evaluate precision. The GEM5000 failed to achieve the required precision for tHb, Hct and iCa, but achieved acceptable precision for all other parameters. Concordance correlation analysis revealed poor correlation for Na, Cl, iCa, K, Hct and tHb, while there was an at least moderate agreement for all other parameters. Passing-Bablok regression revealed significant constant bias for pCO, pO, Cl, and iCa and significant proportional bias for pCO, iCa and SO. Bland-Altman analysis revealed significant systematic bias for Na, Cl, iCa, K, Hct, tHb and SO. This study shows that while precision of the GEM5000 is good, values should not be used interchangeably with data obtained from other blood gas analyzers.
在马的临床实践中会使用不同的血气分析仪。每台机器都需要进行验证,因为它们并非设计用于马匹。本研究的目的是将新上市的GEM5000机器与先前已验证的epoc机器用于马匹血气分析进行比较。在这项前瞻性、非盲法、比较实验室分析仪研究中,对43份马血样本在两台分析仪上进行分析,并通过林氏一致性分析、Passing-Bablok回归分析和Bland-Altman图比较两台机器之间的值。在GEM5000机器上进行重复测量以评估精密度。GEM5000在总血红蛋白(tHb)、血细胞比容(Hct)和离子钙(iCa)方面未能达到所需的精密度,但在所有其他参数方面达到了可接受的精密度。一致性相关分析显示,钠(Na)、氯(Cl)、iCa、钾(K)、Hct和tHb的相关性较差,而所有其他参数至少有中度一致性。Passing-Bablok回归显示,二氧化碳分压(pCO₂)、氧分压(pO₂)、Cl和iCa存在显著的恒定偏差,pCO₂、iCa和血氧饱和度(SO₂)存在显著的比例偏差。Bland-Altman分析显示,Na、Cl、iCa、K、Hct、tHb和SO₂存在显著的系统偏差。本研究表明,虽然GEM5000的精密度良好,但所得值不应与从其他血气分析仪获得的数据互换使用。