牛血液中维生素A浓度的验证:新型现场检测(iCheck™ FLUORO)与高效液相色谱法(HPLC)的比较

Validation of blood vitamin A concentrations in cattle: comparison of a new cow-side test (iCheck™ FLUORO) with high-performance liquid chromatography (HPLC).

作者信息

Raila Jens, Kawashima Chiho, Sauerwein Helga, Hülsmann Nadine, Knorr Christoph, Myamoto Akio, Schweigert Florian J

机构信息

Institute of Nutritional Science, University of Potsdam, Arthur-Scheunert-Allee 114-116, 14558, Nuthetal, Germany.

Obihiro University of Agriculture and Veterinary Medicine, Obihiro, Hokkaido, 080-8555, Japan.

出版信息

BMC Vet Res. 2017 May 10;13(1):126. doi: 10.1186/s12917-017-1042-3.

Abstract

BACKGROUND

Plasma concentration of retinol is an accepted indicator to assess the vitamin A (retinol) status in cattle. However, the determination of vitamin A requires a time consuming multi-step procedure, which needs specific equipment to perform extraction, centrifugation or saponification prior to high-performance liquid chromatography (HPLC).

METHODS

The concentrations of retinol in whole blood (n = 10), plasma (n = 132) and serum (n = 61) were measured by a new rapid cow-side test (iCheck™ FLUORO) and compared with those by HPLC in two independent laboratories in Germany (DE) and Japan (JP).

RESULTS

Retinol concentrations in plasma ranged from 0.033 to 0.532 mg/L, and in serum from 0.043 to 0.360 mg/L (HPLC method). No significant differences in retinol levels were observed between the new rapid cow-side test and HPLC performed in different laboratories (HPLC vs. iCheck™ FLUORO: 0.320 ± 0.047 mg/L vs. 0.333 ± 0.044 mg/L, and 0.240 ± 0.096 mg/L vs. 0.241 ± 0.069 mg/L, lab DE and lab JP, respectively). A similar comparability was observed when whole blood was used (HPLC vs. iCheck™ FLUORO: 0.353 ± 0.084 mg/L vs. 0.341 ± 0.064 mg/L). Results showed a good agreement between both methods based on correlation coefficients of r = 0.87 (P < 0.001) and Bland-Altman blots revealed no significant bias for all comparison.

CONCLUSIONS

With the new rapid cow-side test (iCheck™ FLUORO) retinol concentrations in cattle can be reliably assessed within a few minutes and directly in the barn using even whole blood without the necessity of prior centrifugation. The ease of the application of the new rapid cow-side test and its portability can improve the diagnostic of vitamin A status and will help to control vitamin A supplementation in specific vitamin A feeding regimes such as used to optimize health status in calves or meat marbling in Japanese Black cattle.

摘要

背景

视黄醇的血浆浓度是评估牛维生素A(视黄醇)状态的公认指标。然而,维生素A的测定需要耗时的多步骤程序,在进行高效液相色谱法(HPLC)之前,需要特定设备进行提取、离心或皂化。

方法

采用一种新的快速现场检测方法(iCheck™ FLUORO)测定全血(n = 10)、血浆(n = 132)和血清(n = 61)中的视黄醇浓度,并与德国(DE)和日本(JP)两个独立实验室采用HPLC法测定的结果进行比较。

结果

血浆中视黄醇浓度范围为0.033至0.532mg/L,血清中视黄醇浓度范围为0.043至0.360mg/L(HPLC法)。在不同实验室进行的新快速现场检测与HPLC法之间,视黄醇水平未观察到显著差异(HPLC法与iCheck™ FLUORO法:分别为0.320±0.047mg/L与0.333±0.044mg/L,以及0.240±0.096mg/L与0.241±0.069mg/L,DE实验室和JP实验室)。使用全血时也观察到了类似的可比性(HPLC法与iCheck™ FLUORO法:0.353±0.084mg/L与0.341±0.064mg/L)。基于r = 0.87的相关系数,两种方法结果显示出良好的一致性(P < 0.001),Bland-Altman图显示所有比较均无显著偏差。

结论

使用新的快速现场检测方法(iCheck™ FLUORO),即使使用全血,无需预先离心,也能在几分钟内在牛舍直接可靠地评估牛的视黄醇浓度。新快速现场检测方法应用简便且便于携带,可改善维生素A状态的诊断,并有助于在特定的维生素A喂养方案中控制维生素A的补充,例如用于优化犊牛健康状况或日本黑牛的大理石花纹。