Lopes Bernardo T, Roberts Cynthia J, Elsheikh Ahmed, Vinciguerra Riccardo, Vinciguerra Paolo, Reisdorf Sven, Berger Stefanie, Koprowski Robert, Ambrósio Renato
Rio de Janeiro Corneal Tomography and Biomechanics Study Group, Rio de Janeiro, RJ, Brazil.
Department of Ophthalmology, Federal University of São Paulo, São Paulo, SP, Brazil.
J Ophthalmol. 2017;2017:8515742. doi: 10.1155/2017/8515742. Epub 2017 Jun 6.
To assess the repeatability and reproducibility of dynamic corneal response parameters measured by the Corvis ST (Oculus, Wetzlar, Germany).
One eye randomly selected from 32 healthy volunteers was examined by the Corvis ST. Three different devices were used in an alternated random order for taking three measurements at each device in each subject. Standard intraocular pressure (IOP), the biomechanical-compensated IOP (bIOP), and DCR parameters were evaluated. The within-subject standard deviation (w) and coefficient of variation (CV) were assessed.
Regarding pressure indices, the w was below 1 mmHg for repeatability (0.98 for IOP and 0.89 for bIOP) and the CV was 6.6% for IOP and 6.1% for bIOP. For reproducibility, the w was around 1 mmHg (1.12 for IOP and 1.05 for bIOP) and the CV was 7.6% for IOP and 7.1% for bIOP. Most of DCR indices presented CV for repeatability below 4%. For reproducibility, the CV of most of the indices were below 6%. The deformation amplitude (DA) ratio in 1 mm and integrated radius were below 4% (1.2% and 3.8%, resp.).
The Corvis ST showed good precision (repeatability and reproducibility) for IOP measurements and for DCR in healthy eyes.
评估Corvis ST(德国韦茨拉尔奥culus公司)测量的动态角膜反应参数的可重复性和再现性。
从32名健康志愿者中随机选择一只眼睛,用Corvis ST进行检查。使用三种不同的设备,以交替随机顺序在每个受试者的每个设备上进行三次测量。评估标准眼压(IOP)、生物力学补偿眼压(bIOP)和DCR参数。评估受试者内标准差(w)和变异系数(CV)。
关于压力指标,重复性方面w低于1 mmHg(IOP为0.98,bIOP为0.89),IOP的CV为6.6%,bIOP的CV为6.1%。对于再现性,w约为1 mmHg(IOP为1.12,bIOP为1.05),IOP的CV为7.6%,bIOP的CV为7.1%。大多数DCR指标的重复性CV低于4%。对于再现性,大多数指标的CV低于6%。1 mm处的变形幅度(DA)比值和积分半径低于4%(分别为1.2%和3.8%)。
Corvis ST在健康眼睛的IOP测量和DCR方面显示出良好的精度(可重复性和再现性)。