Jonnal Ravi S, Gorczynska Iwona, Migacz Justin V, Azimipour Mehdi, Zawadzki Robert J, Werner John S
Vision Science and Advanced Retinal Imaging Laboratory, University of California, Davis, Sacramento, California, United States.
Faculty of Physics, Astronomy and Informatics, Nicolaus Copernicus University, Torun, Poland.
Invest Ophthalmol Vis Sci. 2017 Sep 1;58(11):4559-4568. doi: 10.1167/iovs.16-21138.
Optical coherence tomography's (OCT) third outer retinal band has been attributed to the zone of interdigitation between RPE cells and cone outer segments. The purpose of this paper is to investigate the structure of this band with adaptive optics (AO)-OCT.
Using AO-OCT, images were obtained from two subjects. Axial structure was characterized by measuring band 3 thickness and separation between bands 2 and 3 in segmented cones. Lateral structure was characterized by correlation of band 3 with band 2 and comparison of their power spectra. Band thickness and separation were also measured in a clinical OCT image of one subject.
Band 3 thickness ranged from 4.3 to 6.4 μm. Band 2 correlations ranged between 0.35 and 0.41 and power spectra of both bands confirmed peak frequencies that agree with histologic density measurements. In clinical images, band 3 thickness was between 14 and 19 μm. Measurements of AO-OCT of interband distance were lower than our corresponding clinical OCT measurements.
Band 3 originates from a structure with axial extent similar to a single surface. Correlation with band 2 suggests an origin within the cone photoreceptor. These two observations indicate that band 3 corresponds predominantly to cone outer segment tips (COST). Conventional OCT may overestimate both the thickness of band 3 and outer segment length.
光学相干断层扫描(OCT)的视网膜外层第三带被认为与视网膜色素上皮(RPE)细胞和视锥细胞外段之间的指状交叉带有关。本文旨在利用自适应光学(AO)-OCT研究该带的结构。
使用AO-OCT从两名受试者获取图像。通过测量分段视锥细胞中第3带的厚度以及第2带和第3带之间的间距来表征轴向结构。通过第3带与第2带的相关性及其功率谱的比较来表征横向结构。还在一名受试者的临床OCT图像中测量了带厚度和间距。
第3带厚度范围为4.3至6.4μm。第2带的相关性在0.35至0.41之间,两条带的功率谱均证实了与组织学密度测量结果一致的峰值频率。在临床图像中,第3带厚度在14至19μm之间。AO-OCT测量的带间距离低于我们相应的临床OCT测量值。
第3带起源于轴向范围类似于单个表面的结构。与第2带的相关性表明其起源于视锥光感受器内。这两个观察结果表明,第3带主要对应于视锥细胞外段尖端(COST)。传统OCT可能会高估第3带的厚度和外段长度。