Suppr超能文献

三级医疗转诊医院儿科年龄组血糖仪分析性能评估。

Assessment of analytical performance of glucose meter in pediatric age group at tertiary care referral hospital.

作者信息

Rojekar Mohit Vijay, Kumawat Vandana, Panot Jayesh, Khedkar Surekha, Adhe-Rojekar Arati

机构信息

Rajiv Gandhi Medical College, Thane, India.

PD Hinduja National Hospital and Research Center, Mumbai, India.

出版信息

J Diabetes Metab Disord. 2017 Sep 13;16:38. doi: 10.1186/s40200-017-0318-2. eCollection 2017.

Abstract

BACKGROUND

Glucometers are the excellent tools for self-monitoring of blood glucose (SMBG). They are important especially in the circumstances where continuous monitoring is mandatory and at decision making levels. Tight glycemic control protocols are important for preventing the ill effects of fluctuating glucose levels. This increases the use of glucometers in various healthcare settings. As technology advances, glucometers are getting better in terms of quality of results. But still some lacunae are there.

METHODS

Present study was conducted in the tertiary care referral hospital. One hundred twenty five patients were recruited from pediatric wards. Bland-Altman plot, Parke error grid and Surveillance error grid analysis were used for comparing results of glucose meter with that of standard laboratory method.

RESULTS

It is found that there is significant difference between the results by two methods. Though minimal but glucose meter results deviate from the results of standard lab method. This will affect the overall patient care especially in emergency conditions.

CONCLUSIONS

This study is the first of its kind as no similar studies have been reported in the pediatric population. For effective use of glucose meter it should give as accurate as possible estimate of actual glucose levels. Results should not only be accurate but also precise without which critical errors may be possible. We recommend that for any glucose meter there should be regular maintenance as well as calibration is to be done. So that agreement with reference laboratory method is maintained and effective medical decisions are made.

摘要

背景

血糖仪是自我血糖监测(SMBG)的优秀工具。它们在必须进行连续监测的情况下以及决策层面尤为重要。严格的血糖控制方案对于预防血糖水平波动的不良影响至关重要。这增加了血糖仪在各种医疗环境中的使用。随着技术的进步,血糖仪在检测结果质量方面越来越好。但仍然存在一些缺陷。

方法

本研究在三级转诊医院进行。从儿科病房招募了125名患者。使用Bland-Altman图、帕克误差网格和监测误差网格分析来比较血糖仪与标准实验室方法的结果。

结果

发现两种方法的结果之间存在显著差异。虽然差异很小,但血糖仪的结果与标准实验室方法的结果存在偏差。这将影响整体患者护理,尤其是在紧急情况下。

结论

本研究尚属首次,因为儿科人群中尚未有类似研究报道。为了有效使用血糖仪,它应该尽可能准确地估计实际血糖水平。结果不仅要准确,而且要精确,否则可能会出现严重误差。我们建议对任何血糖仪都应进行定期维护以及校准。以便与参考实验室方法保持一致并做出有效的医疗决策。