Magn Reson Med. 2018 Jul;80(1):272-278. doi: 10.1002/mrm.26994. Epub 2017 Nov 6.
Transit delay is a potential source of error in cardiac arterial spin-labeled (ASL) in heart failure or with collateral circulation. This study demonstrates the feasibility of using transit delay insensitive velocity selective ASL and compares its performance with flow-sensitive alternating inversion recovery (FAIR) ASL.
Velocity selective labeling was achieved using an adiabatic BIR8 preparation. FAIR and velocity-selective ASL (VSASL) with various velocity cutoffs (V = 10-40 cm/s) and labeling directions (anterior-posterior X, lateral-septal Y, and apical-basal Z) were carried out in 10 healthy volunteers (1F/9M age 23-30 y). Myocardial blood flow (MBF) and temporal signal-to-noise (TSNR) were measured.
VSASL sensitivity to perfusion decreased with increasing V . At low V (<5 cm/s), spurious labeling of myocardium occurs and overestimates MBF. MBF measured with FAIR (1.12 ± 0.26 ml/g/min) and VASL (1.26 ± 0.27 ml/g/min) at V of 10 cm/s in Z were comparable (TOST with difference of 0.30 ml/g/min, P = 0.049). TSNR was 2.8 times larger using FAIR (13.62 ± 5.25) than in VSASL (4.87 ± 1.58). VSASL was insensitive to perfusion in the Y direction. X and Z performed similarly with TSNR of 4.17 ± 2.32 and 3.97 ± 0.56, respectively.
VSASL is a promising alternative to FAIR ASL in the heart and is well suited for scenarios when transit delays are long. Magn Reson Med 80:272-278, 2018. © 2017 International Society for Magnetic Resonance in Medicine.
在心力衰竭或存在侧支循环的情况下,传输延迟是心脏动脉自旋标记(ASL)中的一个潜在误差源。本研究旨在展示使用无传输延迟敏感速度选择 ASL 的可行性,并比较其与流动敏感交替反转恢复(FAIR)ASL 的性能。
采用绝热 BIR8 制备实现速度选择标记。在 10 名健康志愿者(1 名女性/9 名男性,年龄 23-30 岁)中进行了 FAIR 和速度选择 ASL(VSASL),各种速度截止值(V=10-40cm/s)和标记方向(前后 X、左右 Y、顶底 Z)。测量心肌血流(MBF)和时间信号到噪声比(TSNR)。
VSASL 对灌注的灵敏度随 V 的增加而降低。在低 V(<5cm/s)下,心肌会出现虚假标记并高估 MBF。在 V 为 10cm/s 时,FAIR(1.12±0.26ml/g/min)和 Z 方向 VSASL(1.26±0.27ml/g/min)测量的 MBF 相当(TOST 差值为 0.30ml/g/min,P=0.049)。FAIR 的 TSNR 比 VSASL(4.87±1.58)大 2.8 倍(13.62±5.25)。VSASL 在 Y 方向对灌注不敏感。X 和 Z 方向的 TSNR 分别为 4.17±2.32 和 3.97±0.56,表现相似。
VSASL 是 FAIR ASL 在心脏中的一种有前途的替代方法,非常适合传输延迟较长的情况。
磁共振医学 80:272-278, 2018。©2017 国际磁共振医学学会。