Department of Ophthalmology, The University of Tokyo, Tokyo, Japan; Department of Ophthalmology, Graduate School of Medical Science, Kitasato University, Kanagawa, Japan.
Department of Ophthalmology, The University of Tokyo, Tokyo, Japan.
Am J Ophthalmol. 2018 Mar;187:1-9. doi: 10.1016/j.ajo.2017.12.002. Epub 2017 Dec 15.
The purpose of the current study was to evaluate the test-retest reproducibility and structure-function relationship of the MP-3 microperimeter, compared against the Humphrey Field Analyzer (HFA).
Design: Reliability and validity study.
Institutional, or clinical practice.
Thirty eyes of 30 primary open-angle glaucoma patients were enrolled.
Visual fields (VF) were measured twice with the MP-3 and HFA instruments, using the 10-2 test grid pattern in both perimeters. Ganglion cell complex (GCC) thickness was measured using optical coherence tomography (OCT). Test-retest reproducibility was assessed using the mean absolute deviation (MAD) measure at all 68 VF test points, and also the intraclass correlation coefficient (ICC) of the repeated VF sensitivities. The structure-function relationship between VF sensitivities (measured with MP-3 or HFA) and GCC thickness (adjusted for the retinal ganglion cell displacement) was analyzed using linear mixed modeling.
Reproducibility and structure-function relationship.
The average measurement duration with the HFA 10-2 was 7 minutes and 6 seconds (7m06s) ± 0m49s (mean ± standard deviation). A significantly (P < .001, paired Wilcoxon test) longer measurement duration was observed for the MP-3 test: 10m29s ± 2m55s. There were no significant differences in MAD and ICC values between HFA (MAD; 0.83 ± 0.69 dB and ICC: 0.89 ± 0.69, mean ± standard deviation) and MP-3 (MAD: 0.65 ± 0.67 dB and ICC: 0.89 ± 0.69). MP-3 VF sensitivities had a stronger structure-function relationship with GCC thickness compared to HFA.
The MP-3 microperimeter has a similar test-retest reproducibility to the HFA but a better structure-function relationship.
本研究旨在评估 MP-3 微视野计与 Humphrey 视野分析仪(HFA)相比的复测重现性和结构-功能关系。
设计:可靠性和有效性研究。
机构或临床实践。
30 名原发性开角型青光眼患者的 30 只眼被纳入研究。
使用 MP-3 和 HFA 仪器,在两个视野计上均使用 10-2 测试网格模式,两次测量视野(VF)。使用光学相干断层扫描(OCT)测量神经节细胞复合体(GCC)厚度。使用所有 68 个 VF 测试点的平均绝对偏差(MAD)测量值以及重复 VF 敏感性的组内相关系数(ICC)评估复测重现性。使用线性混合建模分析 VF 敏感性(使用 MP-3 或 HFA 测量)和 GCC 厚度(校正视网膜神经节细胞位移)之间的结构-功能关系。
重现性和结构-功能关系。
HFA 10-2 的平均测量时间为 7 分钟 6 秒(7m06s)±0m49s(平均值±标准差)。MP-3 测试的测量时间明显更长(P<0.001,配对 Wilcoxon 检验):10 分钟 29 秒±2 分钟 55 秒。HFA(MAD:0.83±0.69dB,ICC:0.89±0.69,平均值±标准差)和 MP-3(MAD:0.65±0.67dB,ICC:0.89±0.69)之间的 MAD 和 ICC 值无显著差异。与 HFA 相比,MP-3 VF 敏感性与 GCC 厚度具有更强的结构-功能关系。
MP-3 微视野计与 HFA 具有相似的复测重现性,但具有更好的结构-功能关系。