Hosseinaee Zohreh, Tan Bingyao, Martinez Adam, Bizheva Kostadinka K
Department of System Design Engineering, University of Waterloo, Waterloo, ON, Canada.
Department of Physics and Astronomy, University of Waterloo, Waterloo, ON, Canada.
Transl Vis Sci Technol. 2018 Aug 24;7(4):18. doi: 10.1167/tvst.7.4.18. eCollection 2018 Aug.
To compare the accuracy of Doppler optical coherence tomography (DOCT) and OCT angiography (OCTA) for measuring retinal blood vessel caliber at different flow rates.
A research-grade 1060-nm OCT system with 3.5-μm axial resolution in retinal tissue and 92,000 A scan/s image acquisition rate was used in this study. DOCT and OCTA measurements were acquired both from a flow phantom and in vivo from retinal blood vessels in six male Brown Norway rats. The total retinal blood flow (TRBF) was modified from baseline to 70% and 20% of baseline by reducing the ocular perfusion pressure (OPP). The retinal blood vessel caliber (RBVC) was measured from OCTA and DOCT images. The caliber measurements were conducted by two separate graders using a custom MATLAB-based image processing algorithm.
The RBVC measured with OCTA and DOCT for normal blood flow rates were not significantly different (56.69 ± 12.17 and 57.17 ± 9.46 μm, = 0.27, respectively). However, significant differences were detected when TRBF was reduced to 70% (55.69 ± 11.56 vs. 50.62 ± 8.85 μm, < 0.01) and 20% (50.29 ± 9.29 vs. 44.88 ± 7.13 μm, < 0.01) of baseline.
Reduced TRBF resulted in inaccuracy of the RBVC measurements with DOCT in both the phantom and animal study. This result suggests that OCTA is a more accurate tool for RBVC evaluation when applied to retinal diseases associated with reduced TRBF, such as glaucoma and diabetic retinopathy.
Results from this study are directly applicable to clinical studies of retinal blood flow measured with OCTA and DOCT.
比较多普勒光学相干断层扫描(DOCT)和光学相干断层扫描血管造影(OCTA)在不同流速下测量视网膜血管管径的准确性。
本研究使用了一台研究级1060纳米OCT系统,其在视网膜组织中的轴向分辨率为3.5微米,图像采集速率为92,000次A扫描/秒。DOCT和OCTA测量均在流动模型以及六只雄性挪威棕色大鼠的视网膜血管活体上进行。通过降低眼灌注压(OPP),将总视网膜血流量(TRBF)从基线改变为基线的70%和20%。从OCTA和DOCT图像中测量视网膜血管管径(RBVC)。管径测量由两名独立的分级人员使用基于MATLAB的定制图像处理算法进行。
对于正常血流速度,用OCTA和DOCT测量的RBVC无显著差异(分别为56.69±12.17和57.17±9.46微米,P = 0.27)。然而,当TRBF降至基线的70%(55.69±11.56与50.62±8.85微米,P < 0.01)和20%(50.29±9.29与44.88±7.13微米,P < 0.01)时,检测到显著差异。
在模型和动物研究中,TRBF降低均导致DOCT测量RBVC不准确。该结果表明,当应用于与TRBF降低相关的视网膜疾病(如青光眼和糖尿病视网膜病变)时,OCTA是评估RBVC更准确的工具。
本研究结果可直接应用于用OCTA和DOCT测量视网膜血流的临床研究。