Tagiling Nashrulhaq, Ab Rashid Raizulnasuha, Azhan Siti Nur Amirah, Dollah Norhayati, Geso Moshi, Rahman Wan Nordiana
Medical Radiation Programme, School of Health Sciences, Universiti Sains Malaysia (Health Campus), Kelantan, Malaysia.
Department of Nuclear Medicine, Radiotherapy and Oncology, Hospital Universiti Sains Malaysia (HUSM), Kelantan, Malaysia.
Heliyon. 2018 Oct 17;4(10):e00864. doi: 10.1016/j.heliyon.2018.e00864. eCollection 2018 Oct.
Proper dosimetry settings are crucial in radiotherapy to ensure accurate radiation dose delivery. This work evaluated scanning parameters as affecting factors in reading the dose-response of EBT2 and EBT3 radiochromic films (RCFs) irradiated with clinical photon and electron beams. The RCFs were digitised using Epson Expression 10000XL flatbed scanner and image analyses of net optical density () were conducted using five scanning parameters i.e. film type, resolution, image bit depth, colour to grayscale transformation and image inversion. The results showed that increasing spatial resolution and deepening colour depth did not improve film sensitivity, while grayscale scanning caused sensitivity reduction below than that detected in the Red-channel. It is also evident that invert and colour negative film type selection negated values, hence unsuitable for scanning RCFs. In conclusion, choosing appropriate scanning parameters are important to maintain preciseness and reproducibility in films dosimetry.
在放射治疗中,正确的剂量测定设置对于确保准确的辐射剂量传递至关重要。这项工作评估了扫描参数作为影响读取临床光子和电子束照射的EBT2和EBT3放射变色胶片(RCF)剂量响应的因素。使用爱普生Expression 10000XL平板扫描仪对RCF进行数字化,并使用五个扫描参数,即胶片类型、分辨率、图像位深度、颜色到灰度转换和图像反转,对净光密度()进行图像分析。结果表明,提高空间分辨率和加深颜色深度并不能提高胶片灵敏度,而灰度扫描导致灵敏度降低,低于在红色通道中检测到的灵敏度。同样明显的是,选择反转和彩色负片类型会使值无效,因此不适合扫描RCF。总之,选择合适的扫描参数对于保持胶片剂量测定的准确性和可重复性很重要。