微泡增强聚焦超声频繁应用于血脑屏障开放的安全性评估。

Safety evaluation of frequent application of microbubble-enhanced focused ultrasound blood-brain-barrier opening.

机构信息

Department of Neurosurgery, Chang Gung Memorial Hospital at Linkou, Taoyuan, 333, Taiwan.

Graduate Institute of Clinical Medical Sciences and School of Traditional Chinese Medicine, Chang Gung University, Taoyuan, 333, Taiwan.

出版信息

Sci Rep. 2018 Dec 7;8(1):17720. doi: 10.1038/s41598-018-35677-w.

Abstract

Focused ultrasound (FUS) with the presence of microbubbles induces blood brain barrier (BBB) opening in targeted areas and facilitates drug delivery. However, recent studies have indicated that FUS-BBB opening with excessive exposure levels may be associated with inflammatory response and cellular/tissue damage. Multiple weekly FUS exposures have been shown to be safe for human subjects. However the effect of more frequent FUS exposures is still unknown. This study examines whether frequent focused ultrasound blood brain barrier opening is associated with aggravated behavioral, histopathologic change or brain tissue damage. Two protocols of focused ultrasound blood brain barrier opening were devised using different microbubble doses (0.15 µl/kg and 0.4 µl/kg). Focused ultrasound exposure at a threshold level of BBB-opening, below-threshold level, or above level for intracerebral hemorrhage were delivered every 2 days. Animal behavioral and physiological changes were examined and recorded. Brain tissue was examined for hemorrhage and apoptosis. Results indicate that frequent exposure of excessive focused ultrasound (1.4 mechanical index) produced minor and short-term behavioral changes despite significant tissue damage, while frequent BBB opening with threshold or below-threshold FUS exposure (0.33-0.8 mechanical index) did not cause behavioral or histological change. Immunofluorescent examination of rat brain tissue indicated that excessive doses of microbubble administration induce an apparent cellular apoptotic response, which may be exacerbated by intracerebral hemorrhage. Experimental results suggest that frequent focused ultrasound blood brain barrier opening with sufficient ultrasound exposure level and a microbubble dose can be safe and pose minimal risk to brain tissue.

摘要

聚焦超声(FUS)联合微泡可在靶向区域诱导血脑屏障(BBB)开放,从而促进药物递送。然而,最近的研究表明,过度暴露于 FUS 可能会引起炎症反应和细胞/组织损伤。多次每周 FUS 暴露对人体是安全的。然而,更频繁的 FUS 暴露的影响仍不清楚。本研究旨在探讨频繁的聚焦超声血脑屏障开放是否与加重的行为、组织病理学改变或脑组织损伤有关。本研究设计了两种不同微泡剂量(0.15 μl/kg 和 0.4 μl/kg)的聚焦超声血脑屏障开放方案。采用阈值水平、低于阈值水平或颅内出血水平以上的阈值水平,每隔 2 天进行一次聚焦超声暴露。观察和记录动物的行为和生理变化。检查脑组织是否有出血和细胞凋亡。结果表明,尽管有明显的组织损伤,但频繁暴露于过度的聚焦超声(1.4 机械指数)会产生轻微且短暂的行为变化,而频繁的 BBB 开放(阈值或低于阈值的 FUS 暴露,0.33-0.8 机械指数)不会引起行为或组织学变化。大鼠脑组织的免疫荧光检查表明,微泡给药的高剂量会引起明显的细胞凋亡反应,而颅内出血可能会加剧这种反应。实验结果表明,在足够的超声暴露水平和微泡剂量下,频繁的聚焦超声血脑屏障开放是安全的,对脑组织的风险最小。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/e516/6286368/9cfedd835b2d/41598_2018_35677_Fig1_HTML.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验