Symons Rolf, Sandfort Veit, Mallek Marissa, Ulzheimer Stefan, Pourmorteza Amir
Radiology and Imaging Sciences - National Institutes of Health Clinical Center, Bethesda, MD, USA.
Department of Imaging & Pathology, University Hospitals Leuven, Leuven, Belgium.
Int J Cardiovasc Imaging. 2019 Apr;35(4):733-739. doi: 10.1007/s10554-018-1499-6. Epub 2019 Jan 11.
To evaluate the performance of photon-counting detector (PCD) computed tomography (CT) for coronary artery calcium (CAC) score imaging at standard and reduced radiation doses compared to conventional energy-integrating detector (EID) CT. A dedicated cardiac CT phantom, ten ex vivo human hearts, and ten asymptomatic volunteers underwent matched EID and PCD CT scans at different dose settings without ECG gating. CAC score, contrast, and contrast-to-noise ratio (CNR) were calculated in the cardiac CT phantom. CAC score accuracy and reproducibility was assessed in the ex vivo hearts. Standard radiation dose (120 kVp, reference mAs = 80) in vivo CAC scans were compared against dose-reduced CAC scans (75% dose reduction; reference mAs = 20) for image quality and CAC score reproducibility. Interstudy agreement was assessed by using intraclass correlation (ICC), linear regression, and Bland-Altman analysis with 95% confidence interval limits of agreement (LOA). Calcium-soft tissue contrast and CNR were significantly higher for the PCD CAC scans in the cardiac CT phantom (all P < 0.01). Ex vivo hearts: CAC score reproducibility was significantly higher for the PCD scans at the lowest dose setting (50 mAs) (P = 0.002); score accuracy was similar for both detector systems at all dose settings. In vivo scans: the agreement between standard dose and low dose CAC score was significantly better for the PCD than for the EID with narrower LOA in Bland-Altman analysis, linear regression slopes closer to 1 (0.96 vs. 0.84), and higher ICC values (0.98 vs. 0.93, respectively). Phantom and in vivo human studies showed PCD may significantly improve CAC score image quality and/or reduce CAC score radiation dose while maintaining diagnostic image quality.
为了评估光子计数探测器(PCD)计算机断层扫描(CT)在标准辐射剂量和降低辐射剂量下与传统能量积分探测器(EID)CT相比用于冠状动脉钙化(CAC)评分成像的性能。一个专用的心脏CT模型、十个离体人心脏和十名无症状志愿者在不同剂量设置下进行了非心电门控的匹配EID和PCD CT扫描。在心脏CT模型中计算CAC评分、对比度和对比噪声比(CNR)。在离体心脏中评估CAC评分的准确性和可重复性。将体内CAC扫描的标准辐射剂量(120 kVp,参考mAs = 80)与降低剂量的CAC扫描(剂量降低75%;参考mAs = 20)进行图像质量和CAC评分可重复性的比较。通过组内相关系数(ICC)、线性回归和Bland-Altman分析评估研究间的一致性,并给出95%置信区间的一致性界限(LOA)。在心脏CT模型中,PCD CAC扫描的钙-软组织对比度和CNR显著更高(所有P < 0.01)。离体心脏:在最低剂量设置(50 mAs)下,PCD扫描的CAC评分可重复性显著更高(P = 0.002);在所有剂量设置下,两种探测器系统的评分准确性相似。体内扫描:在Bland-Altman分析中,PCD的标准剂量和低剂量CAC评分之间的一致性明显优于EID,其LOA更窄,线性回归斜率更接近1(分别为0.96对0.84),ICC值更高(分别为0.98对0.93)。模型和体内人体研究表明,PCD在保持诊断图像质量的同时,可能显著提高CAC评分图像质量和/或降低CAC评分辐射剂量。