Impedimetric array in polymer microfluidic cartridge for low cost point-of-care diagnostics.

机构信息

Healthcare Innovation Centre, Teesside University, Middlesbrough, Tees Valley TS1 3BX, UK.

Healthcare Innovation Centre, Teesside University, Middlesbrough, Tees Valley TS1 3BX, UK.

出版信息

Biosens Bioelectron. 2019 Mar 15;129:147-154. doi: 10.1016/j.bios.2018.12.054. Epub 2019 Jan 11.

Abstract

Deep Vein Thrombosis and pulmonary embolism (DVT/PE) is one of the most common causes of unexpected death for hospital in-patients. D-dimer is used as a biomarker within blood for the diagnosis of DVT/PE. We report a low-cost microfluidic device with a conveniently biofunctionalised interdigitated electrode (IDE) array and a portable impedimetric reader as a point-of-care (POC) device for the detection of D-dimer to aid diagnosis of DVT/PE. The IDE array elements, fabricated on a polyethylenenaphtalate (PEN) substrate, are biofunctionalised in situ after assembly of the microfluidic device by electropolymerisation of a copolymer of polypyrrole to which is immobilised a histidine tag anti-D-Dimer antibody. The most consistent copolymer films were produced using chronopotentiometry with an applied current of 5μA for a period of 50 s using a two-electrode system. The quality of the biofunctionalisation was monitored using optical microscopy, chronopotentiometry curves and impedimetric analysis. Measurement of clinical plasma sample with a D-dimer at concentration of 437 ng/mL with 15 biofunctionalised IDE array electrodes gave a ratiometric percentage of sample reading against the blank with an average value of 124 ± 15 at 95% confidence. We have demonstrated the concept of a low cost disposable microfluidic device with a receptor functionalised on the IDE array for impedimetric detection towards POC diagnostics. Changing the receptor on the IDE array would allow this approach to be used for the direct detection of a wide range of analytes in a low cost manner.

摘要

深静脉血栓形成和肺栓塞(DVT/PE)是住院患者中最常见的意外死亡原因之一。D-二聚体作为血液中的生物标志物,用于诊断 DVT/PE。我们报告了一种低成本的微流控设备,该设备具有方便的生物功能化叉指电极(IDE)阵列和便携式阻抗读数器,可作为即时护理(POC)设备,用于检测 D-二聚体,以辅助 DVT/PE 的诊断。IDE 阵列元件在聚萘二甲酸乙二醇酯(PEN)衬底上制造,在微流控设备组装后通过聚吡咯共聚物的电聚合原位生物功能化,其中固定有组氨酸标记的抗 D-二聚体抗体。使用恒电位计时,使用 5μA 的施加电流在双电极系统中进行 50s 的时间,产生了最一致的共聚物膜。使用光学显微镜、恒电位计时曲线和阻抗分析监测生物功能化的质量。使用具有 437ng/mL D-二聚体浓度的临床血浆样本进行测量,使用 15 个生物功能化 IDE 阵列电极获得了相对于空白的样本读数的相对百分比,平均值为 124±15,置信度为 95%。我们已经证明了具有在 IDE 阵列上功能化受体的低成本一次性微流控设备的概念,用于针对 POC 诊断的阻抗检测。改变 IDE 阵列上的受体将允许以低成本的方式直接检测广泛的分析物。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验