Ghazikhanlou-Sani Karim, Rahimi Azizollah, Poorkaveh Maryam, Eynali Samira, Koosha Fereshteh, Shoja Mohsen
Department of Radiology, Paramedical School, Hamadan University of Medical Sciences, Hamadan, Iran.
Faculty of Medicine, Department of Medical Physics, Ahvaz Judishapur University of Medical Sciences, Ahvaz, Iran.
Interv Med Appl Sci. 2018 Sep;10(3):121-126. doi: 10.1556/1646.10.2018.18.
Operating rooms in hospitals are facilitated with different types of electronic systems, which produce electromagnetic waves. High intensities of magnetic waves may have harmful effects on biological environments. This study aims to evaluate the electromagnetic field intensity at different parts of operating rooms at the first stage and estimate the occupational exposure to operating room personnel at the next phase.
At this cross-sectional study, the magnetic field intensity was evaluated using teslameter at several parts of operating rooms, during operating procedures, while electrical instruments were working. Background electromagnetic field intensity was measured when all the electrical systems were idle. Statistical analysis was performed using SPSS software. The results were compared with ICNIRP standards.
The maximum intensity of magnetic field was measured around high-voltage systems at the distance of 50 cm in the personnel's standing area at DCR and PCNL operating procedures were 5.9 and 5.6, respectively. The number of on-mode electrical systems was inconsistent with the intensity of electromagnetic fields at the standing area of operating room personnel's. The intensity of magnetic fields around high-voltage systems, which was about 46.75 mG at the distance of 10 cm, was the highest among measured electromagnetic fields.
The highest magnetic field intensity measured in this study was related to high-voltage systems and is lower than advised intensity by ICNIRP for occupational exposure. Based on this study, it can be concluded that there are no considerable risks of electromagnetic exposure for operating room personnels.
医院手术室配备了不同类型的电子系统,这些系统会产生电磁波。高强度的电磁波可能会对生物环境产生有害影响。本研究旨在第一阶段评估手术室不同部位的电磁场强度,并在下一阶段估算手术室工作人员的职业暴露情况。
在这项横断面研究中,在手术过程中、电气设备运行时,使用特斯拉计在手术室的几个部位评估磁场强度。当所有电气系统闲置时,测量背景电磁场强度。使用SPSS软件进行统计分析。将结果与国际非电离辐射防护委员会(ICNIRP)的标准进行比较。
在DCR和PCNL手术过程中,人员站立区域距离高压系统50厘米处测得的磁场最大强度分别为5.9和5.6。开启模式的电气系统数量与手术室人员站立区域的电磁场强度不一致。在测得的电磁场中,高压系统周围距离10厘米处的磁场强度最高,约为46.75毫高斯。
本研究中测得的最高磁场强度与高压系统有关,且低于ICNIRP建议的职业暴露强度。基于本研究,可以得出结论,手术室人员不存在相当大的电磁暴露风险。