1 Department of Diagnostic Radiology & Nuclear Medicine, Hamamatsu University School of Medicine, Hamamatsu, Shizuoka, Japan.
2 Central Research Laboratory, Hamamatsu Photonics K.K., Hamamatsu, Shizuoka, Japan.
Technol Cancer Res Treat. 2019 Jan 1;18:1533033819830411. doi: 10.1177/1533033819830411.
We measured total hemoglobin concentrations in breast tumors by near-infrared time-resolved spectroscopy. Muscles interfere with measurement when the probe is close to the chest wall. Since the target area of measurement depends on the distance between the light source and probe detector, we inferred that this issue could be solved by reducing the source-detector distance. The purpose of this study was to examine the effects of the source-detector distance on the measurement of total hemoglobin concentration in the breast. We examined 26 patients with breast tumors. Total hemoglobin concentration was measured in tumors and the contralateral normal breasts at source-detector distances of 20 and 30 mm. The difference in total hemoglobin concentration between each tumor and the contralateral breast was calculated. The normal breast total hemoglobin concentration was significantly smaller for the source-detector distance of 20 mm than for the source-detector distance of 30 mm. Differences in source-detector distance did not significantly affect tumor total hemoglobin. The difference in total hemoglobin concentration between the tumor and the contralateral breast obtained at the source-detector distance of 20 mm was significantly higher than that obtained at the source-detector distance of 30 mm. From these results, we considered that measurement with a source-detector distance of 20 mm is less affected by the chest wall than with a source-detector distance of 30 mm and that the difference in total hemoglobin concentration between the tumor and the contralateral breast at a source-detector distance of 20 mm can better reflect the net total hemoglobin concentrations of the breast tumors. In conclusion, using a probe with a source-detector distance of 20 mm can more accurately evaluate the total hemoglobin concentration in breast tumors.
我们通过近红外时间分辨光谱法测量乳腺肿瘤中的总血红蛋白浓度。当探头靠近胸壁时,肌肉会干扰测量。由于测量的目标区域取决于光源和探头探测器之间的距离,我们推断通过减小源探测器距离可以解决这个问题。本研究旨在探讨源探测器距离对乳腺总血红蛋白浓度测量的影响。我们检查了 26 名患有乳腺肿瘤的患者。在源探测器距离为 20 和 30 毫米时,测量肿瘤和对侧正常乳房中的总血红蛋白浓度。计算每个肿瘤与对侧乳房之间的总血红蛋白浓度差异。源探测器距离为 20 毫米时,正常乳房的总血红蛋白浓度明显小于源探测器距离为 30 毫米时。源探测器距离的差异对肿瘤总血红蛋白没有显著影响。源探测器距离为 20 毫米时获得的肿瘤和对侧乳房之间的总血红蛋白浓度差异明显高于源探测器距离为 30 毫米时。从这些结果中,我们认为与源探测器距离为 30 毫米相比,源探测器距离为 20 毫米的测量受胸壁的影响更小,并且源探测器距离为 20 毫米时肿瘤和对侧乳房之间的总血红蛋白浓度差异可以更好地反映乳腺肿瘤的总血红蛋白浓度。总之,使用源探测器距离为 20 毫米的探头可以更准确地评估乳腺肿瘤中的总血红蛋白浓度。