Lesiuk Mark J, Spencer David P, Chan Alexander K, Voroney Jon-Paul, Lau Harold
Department of Radiation Therapy, Tom Baker Cancer Centre, Calgary, Canada.
Department of Medical Physics, Tom Baker Cancer Centre, Calgary, Canada; University of Calgary, Calgary, Canada.
J Med Imaging Radiat Sci. 2012 Dec;43(4):239-244. doi: 10.1016/j.jmir.2012.07.001. Epub 2012 Sep 24.
To measure setup and intrafraction variability for intracranial targets during treatment of patients immobilized with a Brainlab, Inc. thermoplastic head mask using ExacTrac© imaging on the treatment unit.
Between November 2007 and June 2008, 12 patients were treated with cranial fractionated stereotactic radiation therapy for 25-28 fractions. Patient offset was calculated by the ExacTrac© algorithm from the planning computed tomography and orthogonal kV images taken at the time of treatment. Shifts of the patient position were done when the initial offset was greater than 1.5 mm. If a shift was made, post-shift images were acquired before treatment. Images were also acquired post-treatment to determine intrafraction movement.
A total of 849 pairs of images were obtained. The averaged results of all 12 patients were: setup offset was 1.98 mm, treatment offset was 0.57 mm, and post-treatment offset was 0.66 mm. Of 303 fractions treated, 30 fractions had a post-treatment offset greater than 1.5 mm.
Quantitative analysis of the data set including variability in patient setup and intrafraction motion shows patient position can be reproduced within 1.0 mm using Brainlab thermoplastic head mask systems using the ExacTrac© system.
使用治疗单元上的ExacTrac©成像技术,测量使用Brainlab公司热塑性头罩固定的患者在颅内靶区治疗期间的摆位和分次治疗内的变异性。
在2007年11月至2008年6月期间,12例患者接受了颅部分次立体定向放射治疗,共25 - 28次分割。通过ExacTrac©算法根据治疗时获取的计划计算机断层扫描和正交千伏图像计算患者偏移量。当初始偏移量大于1.5毫米时,进行患者位置的移位。如果进行了移位,则在治疗前获取移位后的图像。治疗后也获取图像以确定分次治疗内的运动情况。
共获得849对图像。所有12例患者的平均结果为:摆位偏移量为1.98毫米,治疗偏移量为0.57毫米,治疗后偏移量为0.66毫米。在303次治疗分割中,有30次治疗后的偏移量大于1.5毫米。
对包括患者摆位变异性和分次治疗内运动情况的数据集进行定量分析表明,使用Brainlab热塑性头罩系统和ExacTrac©系统,患者位置可在1.0毫米内重复。