Hoffner S E
Department of Bacteriology, National Bacteriological Laboratory, Stockholm, Sweden.
Diagn Microbiol Infect Dis. 1988 May;10(1):1-6. doi: 10.1016/0732-8893(88)90121-6.
A reconsideration of the laboratory methods used for primary isolation of mycobacteria other than Mycobacterium tuberculosis is needed due to the increasingly recognized importance of such mycobacterial infections in immunocompromised patients. One example of this is the severe opportunistic infections caused by Mycobacterium avium complex among AIDS patients. In this study, the Bactec radiometric system was compared to conventional culture on solid medium for the detection of M. avium complex in 3,612 selected clinical specimens, mainly of extrapulmonary origin. Of a total number of 63 M. avium complex isolates, the Bactec system detected 58 (92%), compared to 37 (59%) for conventional culture. A much more rapid detection was attained with radiometric technique than with conventional culture. The mean detection time for the cultures positive with both methods was 7.1 and 28.3 days, respectively. The Bactec radiometric system achieves a rapid and significantly more sensitive detection and seems to be an excellent complement to conventional culture in the laboratory diagnosis of infections with the M. avium complex.
由于免疫功能低下患者中此类分枝杆菌感染的重要性日益得到认可,因此需要重新考虑用于结核分枝杆菌以外的分枝杆菌初次分离的实验室方法。其中一个例子是艾滋病患者中由鸟分枝杆菌复合体引起的严重机会性感染。在本研究中,将Bactec放射测量系统与固体培养基上的传统培养方法进行比较,以检测3612份主要来自肺外的选定临床标本中的鸟分枝杆菌复合体。在总共63株鸟分枝杆菌复合体分离株中,Bactec系统检测到58株(92%),而传统培养法检测到37株(59%)。与传统培养法相比,放射测量技术的检测速度要快得多。两种方法均呈阳性的培养物的平均检测时间分别为7.1天和28.3天。Bactec放射测量系统实现了快速且明显更灵敏的检测,似乎是鸟分枝杆菌复合体感染实验室诊断中传统培养的极佳补充。