NEMA Performance Evaluation of CareMiBrain dedicated brain PET and Comparison with the whole-body and dedicated brain PET systems.

机构信息

Instituto de Instrumentación para la Imagen Molecular (i3M), Universitat Politècnica de València - CSIC, Valencia, Spain.

Instituto de Automática e Informática Industrial (ai2), Universitat Politècnica de València, Valencia, Spain.

出版信息

Sci Rep. 2019 Oct 29;9(1):15484. doi: 10.1038/s41598-019-51898-z.

Abstract

This article presents system performance studies of the CareMiBrain dedicated brain PET according to NEMA NU 2-2012 (for whole-body PETs) and NU 4-2008 (for preclinical PETs). This scanner is based on monolithic LYSO crystals coupled to silicon photomultipliers. The results obtained for both protocols are compared with current commercial whole body PETs and dedicated brain PETs found in the literature. Spatial resolution, sensitivity, NECR and scatter-fraction are characterized with NEMA standards, as well as an image quality study. A customized image quality phantom is proposed as NEMA phantoms do not fulfil the necessities of dedicated brain PETs. The full-width half maximum of the radial/tangential/axial spatial resolution of CareMiBrain reconstructed with FBP at 10 and 100 mm from the system center were, respectively, 1.87/1.68/1.39 mm and 1.86/1.91/1.40 mm (NU 2-2012) and 1.58/1.45/1.40 mm and 1.64/1.66/1.44 mm (NU 4-2008). Peak NECR was 49 kcps@287 MBq with a scatter fraction of 48% using NU 2-2012 phantom. The sensitivity was 13.82 cps/kBq at the center of the FOV (NU 2-2012) and 10% (NU 4-2008). Contrast recovery coefficients for customizing image quality phantom were 0.73/0.78/1.14/1.01 for the 4.5/6/9/12 mm diameter rods. The performance characteristics of CareMiBrain are at the top of the current technologies for PET systems. Dedicated brain PET systems significantly improve spatial resolution and sensitivity, but present worse results in count rate measurements and scatter-fraction tests. As for the comparison of preclinical and clinical standards, the results obtained for solid and liquid sources were similar.

摘要

本文根据 NEMA NU 2-2012(用于全身 PET)和 NU 4-2008(用于临床前 PET)标准,对 CareMiBrain 专用脑 PET 的系统性能进行了研究。该扫描仪基于与硅光电倍增管耦合的块状 LYSO 晶体。根据 NEMA 标准,对这两种方案获得的结果与文献中现有的商用全身 PET 和专用脑 PET 进行了比较。空间分辨率、灵敏度、NECR 和散射分数用 NEMA 标准进行了表征,并进行了图像质量研究。提出了一种定制的图像质量体模,因为 NEMA 体模不能满足专用脑 PET 的需要。CareMiBrain 系统中心 10 和 100mm 处 FBP 重建的径向/切向/轴向空间分辨率的全宽半最大值分别为 1.87/1.68/1.39mm 和 1.86/1.91/1.40mm(NU 2-2012)和 1.58/1.45/1.40mm 和 1.64/1.66/1.44mm(NU 4-2008)。使用 NU 2-2012 体模,峰值 NECR 为 49 kcps@287 MBq,散射分数为 48%。在 FOV 中心的灵敏度为 13.82cps/kBq(NU 2-2012)和 10%(NU 4-2008)。定制图像质量体模的对比恢复系数对于 4.5、6、9 和 12mm 直径的棒,分别为 0.73/0.78/1.14/1.01。CareMiBrain 的性能特点在 PET 系统的当前技术中处于领先地位。专用脑 PET 系统显著提高了空间分辨率和灵敏度,但在计数率测量和散射分数测试中表现较差。至于临床前和临床标准的比较,固体和液体源的结果相似。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/4875/6820763/4ca1bf568cc6/41598_2019_51898_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验