附着体粘结方案对矫治器治疗中附着体精度的影响。

Influence of attachment bonding protocol on precision of the attachment in aligner treatments.

作者信息

Weckmann Julia, Scharf Sven, Graf Isabelle, Schwarze Jörg, Keilig Ludger, Bourauel Christoph, Braumann Bert

机构信息

Orthodontic practice, Mittelstraße 4, 53879, Euskirchen, Germany.

Department of Orthodontics, University Hospital of Cologne, Kerpener Straße 32, 50931, Cologne, Germany.

出版信息

J Orofac Orthop. 2020 Jan;81(1):30-40. doi: 10.1007/s00056-019-00204-7. Epub 2019 Dec 13.

Abstract

AIM

Precise bonding of attachments in aligner treatment is crucial to achieve the intended tooth movement. Thus, we evaluated five different bonding protocols for attachments used in aligner treatments with the goal of identifying the most precise protocol.

METHODS

One ellipsoid and one rectangular attachment were exemplarily chosen and examined. All attachments were bonded using the same template aligner, which was produced by thermoforming. The bonding process was repeated 30 times with each protocol for statistical analysis. The protocols differed in the type of composite used (high viscous, low viscous), the additional application of a perforation in the attachment reservoir, and the use of a two-phase procedure with high viscous composite. All bonded attachments were laser scanned and compared to the attachments of a master model using a surface/surface matching algorithm to determine the precision, i.e., highest aberrations (maximum vertical distance) between the attachment of the master and results obtained from the various bonding protocols. Furthermore, the excessive amounts of composite bonded around the attachments were measured and compared.

RESULTS

The two-phase procedure had a median aberration of 0.13 mm and a median amount of excessive composite of 7.40 mm (ellipsoid attachment). With the low viscous composite without a perforation in the template reservoir, similar results were achieved (median 0.13 mm) regarding the maximum vertical distance, but had an excess area of 33.50 mm.

CONCLUSION

In this in vitro study, the bonding protocol influenced the precision of bonded attachments. The bonding protocol with high viscous composite without a perforation in the attachment reservoir was the most inaccurate. The use of a low viscous composite or attachments made by a two-phase procedure with high viscous composite revealed more precise results.

摘要

目的

在矫治器治疗中,附件的精确粘接对于实现预期的牙齿移动至关重要。因此,我们评估了用于矫治器治疗的附件的五种不同粘接方案,目的是确定最精确的方案。

方法

示例性地选择并检查了一个椭圆形附件和一个矩形附件。所有附件均使用通过热成型制作的相同模板矫治器进行粘接。每个方案的粘接过程重复30次以进行统计分析。各方案在所用复合材料类型(高粘性、低粘性)、在附件储存器中额外打孔以及使用高粘性复合材料的两步法方面存在差异。对所有粘接的附件进行激光扫描,并使用表面/表面匹配算法与主模型的附件进行比较,以确定精度,即主模型附件与各种粘接方案所得结果之间的最大像差(最大垂直距离)。此外,测量并比较了附件周围粘接的多余复合材料量。

结果

两步法的中位像差为0.13毫米,多余复合材料的中位量为7.40毫米(椭圆形附件)。对于最大垂直距离,在模板储存器中不打孔的低粘性复合材料也取得了类似结果(中位值0.13毫米),但其多余面积为33.50毫米。

结论

在这项体外研究中,粘接方案影响了粘接附件的精度。在附件储存器中不打孔的高粘性复合材料粘接方案最不准确。使用低粘性复合材料或采用高粘性复合材料的两步法制作的附件显示出更精确的结果。