Marroquín Benjamín Briseño, Fernández Claudia Cortazar, Schmidtmann Irene, Willershausen Brita, Goldberg Fernando
Department of Operative Dentistry, Johannes Gutenberg University Medical Center, Mainz, Germany.
Poliklinik für Zahnerhaltung, Langenbeckstr. 1, Mainz, 55131, Germany.
Head Face Med. 2014 Dec 23;10:57. doi: 10.1186/s13005-014-0057-2.
The detection of possible root canal perforations caused during a metal post placement is frequently difficult to diagnose. The aim of this study was to investigate the accuracy of apex locators to diagnose such perforations.
Thirty mesiolingual root canals of extracted mandibular molars were instrumented (30/.04) and a post space was prepared. A root canal perforation was intentionally made at the bi-furcation (n = 15). Metal posts were cemented in 15 perforated and 15 non-perforated root canals. The teeth roots were embedded in an agar-agar solution. The resulting measurements ("short" or "beyond" the apex) disclosed if a perforation could be identified with five different apex locators (ProPex II, Elements Apex Locator, Apex NRG, Raypex 5 and Raypex 6). The sensitivity and specificity (95% interval confidence) were calculated.
All devices excluded the absence of perforations (100% with 95% confidence interval [78%; 100%] specificity). The Apex NRG and Raypex 6 detected all perforations (100% with 95% confidence interval [78%; 100%] sensitivity). The ProPex II, Elements Apex Locator, and Raypex 5 detected 14 out of 15 perforations (93% with 95% confidence interval [68%; 100%] sensitivity).
All devices determined root canal perforations, due to metallic posts, within clinical acceptable ranges.
在放置金属桩过程中导致的可能的根管穿孔常常难以诊断。本研究的目的是调查根尖定位仪诊断此类穿孔的准确性。
对30颗拔除的下颌磨牙的近中舌根管进行预备(30/.04)并制备桩道。在根分叉处故意造成15个根管穿孔。将金属桩粘结在15个有穿孔和15个无穿孔的根管中。将牙根嵌入琼脂溶液中。通过测量结果(“根尖内”或“根尖外”)来判断是否能用五种不同的根尖定位仪(ProPex II、Elements Apex Locator、Apex NRG、Raypex 5和Raypex 6)识别穿孔。计算敏感性和特异性(95%置信区间)。
所有设备均能排除无穿孔情况(特异性为100%,95%置信区间为[78%;100%])。Apex NRG和Raypex 6能检测到所有穿孔(敏感性为100%,95%置信区间为[78%;100%])。ProPex II、Elements Apex Locator和Raypex 5能检测到15个穿孔中的14个(敏感性为93%,95%置信区间为[68%;100%])。
所有设备在临床可接受范围内均能确定因金属桩导致的根管穿孔。