Nakakura Shunsuke, Mori Etsuko, Fujio Yuki, Fujisawa Yasuko, Matsuya Kanae, Kobayashi Yui, Oogi Satomi, Shiraishi Miku, Tabuchi Hitoshi, Asaoka Ryo, Kiuchi Yoshiaki
Department of Ophthalmology, Saneikai Tsukazaki Hospital, Himeji, Japan.
Department of Ophthalmology, The University of Tokyo, Tokyo, Japan.
Optom Vis Sci. 2020 Feb;97(2):128-133. doi: 10.1097/OPX.0000000000001472.
This study is the first to show that the manual upper eyelid elevation (manual UEE) that is commonly used to prevent disruption of the IOP measurement due to blinking or upper eyelid contact with the tip of the tonometer does not affect the IOP values.
We investigated whether manual UEE affects the IOP readings using three rebound tonometers (Icare TA01i, Icare PRO, and Icare ic100) and Goldmann applanation tonometry (GAT).
One eye was measured for 101 patients (56 eyes of primary open-angle glaucoma patients and 45 healthy subjects). The IOPs were measured without and with manual UEE. Each IOP was measured twice; the measurement order using the tonometers was randomly selected. In addition, palpebral fissure height (distance between the upper and lower eyelids) was measured.
The IOPs without manual UEE were 12.1 ± 2.9, 13.3 ± 2.7, 11.7 ± 2.9, and 16.0 ± 3.2 mmHg (Icare TA01i, Icare PRO, Icare ic100, and GAT), and those with manual UEE were 12.3 ± 3.0, 13.3 ± 2.8, 11.7 ± 2.9, and 16.0 ± 3.3, respectively. No significant difference was found between the IOP without and with manual UEE (IOP difference; all, P > .50; paired t test). Multiple linear regression analyses revealed that palpebral fissure height did not affect IOP difference for any of the tonometers.
Simple manual UEE when measuring the IOP has little effect on the IOP obtained using all current rebound tonometers and GAT.
本研究首次表明,常用于防止因眨眼或上眼睑接触眼压计尖端而干扰眼压测量的手动上睑提升(manual UEE)不会影响眼压值。
我们使用三种回弹式眼压计(Icare TA01i、Icare PRO和Icare ic100)和Goldmann压平眼压计(GAT)研究了手动UEE是否会影响眼压读数。
对101例患者(56例原发性开角型青光眼患者的56只眼和45名健康受试者)的一只眼睛进行测量。在不进行和进行手动UEE的情况下测量眼压。每个眼压测量两次;使用眼压计的测量顺序是随机选择的。此外,测量睑裂高度(上下眼睑之间的距离)。
不进行手动UEE时的眼压分别为12.1±2.9、13.3±2.7、11.7±2.9和16.0±3.2 mmHg(Icare TA01i、Icare PRO、Icare ic100和GAT),进行手动UEE时的眼压分别为12.3±3.0、13.3±2.8、11.7±2.9和16.0±3.3 mmHg。不进行和进行手动UEE时的眼压之间未发现显著差异(眼压差异;所有,P>.50;配对t检验)。多元线性回归分析显示,睑裂高度对任何眼压计的眼压差异均无影响。
测量眼压时简单的手动UEE对使用所有当前回弹式眼压计和GAT获得的眼压影响很小。