Hiroshima High-Precision Radiotherapy Cancer Center, 3-2-2, Futabanosato, Higashi-ku, Hiroshima, 732-0057, Japan.
Department of Radiation Oncology, Graduate School of Biomedical & Health Sciences, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima, 734-8551, Japan.
Med Phys. 2020 Apr;47(4):1509-1522. doi: 10.1002/mp.14077. Epub 2020 Feb 29.
In photon radiation therapy, computed tomography (CT) numbers are converted into values for mass density (MD) or relative electron density to water (RED). CT-MD or CT-RED calibration tables are relevant for human body dose calculation in an inhomogeneous medium. CT-MD or CT-RED calibration tables are influenced by patient imaging (CT scanner manufacturer, scanning parameters, and patient size), the calibration process (tissue-equivalent phantom manufacturer, and selection of tissue-equivalent material), differences between tissue-equivalent materials and standard tissues, and the dose calculation algorithm applied; however, a CT number calibration audit has not been established. The purposes of this study were to develop a postal audit phantom, and to establish a CT number calibration audit process.
A conventional stoichiometric calibration conducts a least square fit of the relationships between the MD, material weight, and measured CT number, using two parameters. In this study, a new stoichiometric CT number calibration scheme has been empirically established, using three parameters to harmonize the calculated CT number with the measured CT number for air and lung tissue. In addition, the suitable material set and the minimal number of materials required for stoichiometric CT number calibration were determined. The MDs and elemental weights from the International Commission on Radiological Protection Publication 110 were used as standard tissue data, to generate the CT-MD and CT-RED calibration tables. A small-sized, CT number calibration phantom was developed for a postal audit, and stoichiometric CT number calibration with the phantom was compared to the CT number calibration tables registered in the radiotherapy treatment planning systems (RTPSs) associated with five radiotherapy institutions.
When a least square fit was performed for the stoichiometric CT number calibration with the three parameters, the calculated CT number showed better agreement with the measured CT number. We established stoichiometric CT number calibration using only two materials because the accuracy of the process was determined not by the number of used materials but by the number of elements contained. The stoichiometric CT number calibration was comparable to the tissue-substitute calibration, with a dose difference less than 1%. An outline of the CT number calibration audit was demonstrated through a multi-institutional study.
We established a new stoichiometric CT number calibration method for validating the CT number calibration tables registered in RTPSs. We also developed a CT number calibration phantom for a postal audit, which was verified by the performances of multiple CT scanners located at several institutions. The new stoichiometric CT number calibration has the advantages of being performed using only two materials, and decreasing the difference between the calculated and measured CT numbers for air and lung tissue. In the future, a postal CT number calibration audit might be achievable using a smaller phantom.
在光子放射治疗中,将计算机断层扫描(CT)的数值转换为质量密度(MD)或相对电子密度对水(RED)。CT-MD 或 CT-RED 校准表与不均匀介质中的人体剂量计算有关。CT-MD 或 CT-RED 校准表受患者成像(CT 扫描仪制造商、扫描参数和患者大小)、校准过程(组织等效体模制造商和组织等效材料的选择)、组织等效材料与标准组织之间的差异以及所应用的剂量计算算法的影响;然而,尚未建立 CT 数校准审核。本研究的目的是开发一种邮政审核体模,并建立 CT 数校准审核流程。
传统的化学计量学校准使用两个参数,对 MD、材料重量和测量的 CT 数之间的关系进行最小二乘拟合。在本研究中,通过使用三个参数来协调空气和肺组织的计算 CT 数与测量 CT 数,已经建立了一种新的化学计量 CT 数校准方案。此外,确定了适合的材料集和化学计量 CT 数校准所需的最少材料数量。国际辐射防护委员会第 110 号出版物中的 MD 和元素重量被用作标准组织数据,以生成 CT-MD 和 CT-RED 校准表。为了进行邮政审核,开发了一个小型 CT 数校准体模,并将其与五个放射治疗机构相关联的放射治疗计划系统(RTPS)中注册的 CT 数校准表进行了化学计量 CT 数校准比较。
当使用三个参数进行化学计量 CT 数校准的最小二乘拟合时,计算的 CT 数与测量的 CT 数吻合更好。由于过程的准确性不是由使用的材料数量决定的,而是由包含的元素数量决定的,因此我们仅使用两种材料建立了化学计量 CT 数校准。化学计量 CT 数校准与组织替代校准相当,剂量差异小于 1%。通过多机构研究展示了 CT 数校准审核的概述。
我们建立了一种新的化学计量 CT 数校准方法,用于验证 RTPS 中注册的 CT 数校准表。我们还开发了一种用于邮政审核的 CT 数校准体模,该体模通过位于多个机构的多个 CT 扫描仪的性能得到了验证。新的化学计量 CT 数校准具有仅使用两种材料的优点,并减少了空气和肺组织的计算和测量 CT 数之间的差异。在未来,使用更小的体模可能可以实现邮政 CT 数校准审核。