新型扫频源眼前节光学相干断层扫描与Scheimpflug 相机测量角膜曲率的比较。

Comparison of a New Swept-Source Anterior Segment Optical Coherence Tomography and a Scheimpflug Camera for Measurement of Corneal Curvature.

机构信息

State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou, China.

Department of Ophthalmology, University of California, San Francisco, CA; and.

出版信息

Cornea. 2020 Jul;39(7):818-822. doi: 10.1097/ICO.0000000000002280.

Abstract

PURPOSE

To evaluate the agreement of corneal curvature measured by a newly introduced swept-source anterior segment optical coherence tomography (CASIA2) and Pentacam (Oculus, Wetzlar, Germany).

METHODS

Corneal curvature was measured in 49 healthy subjects (age: 24.8 ± 4.36) by using CASIA2 and Pentacam. Repeatability was evaluated for CASIA2, and agreement between the CASIA2 and Pentacam was assessed. The coefficient of repeatability, the relative coefficient of repeatability, the range of the limits of agreement were analyzed, and Bland-Altman plots were performed to assess the interdevice agreement of measurement.

RESULTS

For the intradevice reliability of the CASIA2, the mean difference between the 2 examinations was from 0.02 to 0.14 D for the corneal curvatures and the limits of agreement range was from 0.45 to 2.42 D for the corneal curvatures; there was no statistically significant difference found between the first and second examinations obtained from CASIA2 (P > 0.05). For agreement between the 2 devices, the Bland-Altman analysis of the posterior corneal curvature showed a mean difference of 0.16 D (P = 0.047) in the flat posterior corneal curvature and 0.16 D (P = 0.046) in the steep posterior corneal curvature. For the measurement of anterior corneal curvature and true net power (TNP) with the Pentacam and CASIA2, there was no significant difference in the flat anterior corneal curvature (P = 0.585), steep anterior corneal curvature (P = 0.274), flat TNP (P = 0.685), and steep TNP (P = 0.447).

CONCLUSIONS

Our study indicated generally good agreement of corneal curvature measurement by using Pentacam and CASIA2. We consequently expect no difference in clinical outcome when measurements for corneal curvature, especially anterior corneal curvature and TNP, are performed with either of the devices.

摘要

目的

评估新型扫频源眼前节光学相干断层扫描仪(CASIA2)和 Pentacam(德国 Oculus,Wetzlar)测量角膜曲率的一致性。

方法

使用 CASIA2 和 Pentacam 对 49 名健康受试者(年龄:24.8±4.36 岁)的角膜曲率进行测量。评估了 CASIA2 的重复性,并评估了 CASIA2 与 Pentacam 的一致性。分析了重复性系数、相对重复性系数、一致性界限范围,并进行了 Bland-Altman 图分析以评估两种设备之间的测量一致性。

结果

对于 CASIA2 的设备内可靠性,两次检查之间角膜曲率的平均差异在 0.02 至 0.14 D 之间,角膜曲率的一致性界限范围在 0.45 至 2.42 D 之间;从 CASIA2 获得的第一次和第二次检查之间没有发现统计学上的显著差异(P>0.05)。对于两种设备之间的一致性,后角膜曲率的 Bland-Altman 分析显示,平坦后角膜曲率的平均差异为 0.16 D(P=0.047),陡峭后角膜曲率的平均差异为 0.16 D(P=0.046)。对于 Pentacam 和 CASIA2 测量前角膜曲率和真实净力(TNP),平坦前角膜曲率(P=0.585)、陡峭前角膜曲率(P=0.274)、平坦 TNP(P=0.685)和陡峭 TNP(P=0.447)之间无显著差异。

结论

我们的研究表明,使用 Pentacam 和 CASIA2 测量角膜曲率具有较好的一致性。因此,当使用这两种设备中的任何一种进行角膜曲率测量,特别是前角膜曲率和 TNP 测量时,我们预计不会对临床结果产生影响。