Determination of cardiac output by single gated pulsed Doppler echocardiography.

作者信息

Fast J H, van den Merkhof L, Blans W, van Leeuwen K, Uijen G

机构信息

Department of Cardiology, Sint Radboud Hospital, University of Nijmegen, The Netherlands.

出版信息

Int J Cardiol. 1988 Oct;21(1):33-42. doi: 10.1016/0167-5273(88)90006-x.

Abstract

The aim of the study was to evaluate the use of single gated pulsed Doppler cross-sectional echocardiography for measurement of the cardiac output. Pulsed Doppler echocardiography was used to assess both area and blood velocity at the aortic orifice. Stroke volume estimates were determined by multiplying area by systolic time velocity integral as measured from the parasternal and apical approach, respectively. We investigated a group of 17 healthy individuals and a group of 20 patients with coronary artery disease. In the first group no change was found in aortic area during systole (P less than 0.01). During a follow-up of two weeks no change in aortic area was observed either (P less than 0.01). Intraobserver, interobserver and day-to-day variability of the aortic area, expressed as the coefficient of variation was 3.6 +- 5.2%, 4.6 +- 5.7% and 7.8 +- 3.8% (mean +- 1 SD), respectively. The interobserver variability of the time velocity integrals was 6.0 +- 6.2%. In the second group the cardiac output as measured with the thermodilution method ranged from 3.8 to 8.0 l/min. Comparison of the Doppler and thermodilution technique for measurement of cardiac output showed a correlation coefficient of r = 0.76 (P less than 0.001) and the following regression equation: CO (Doppler) = 1.0 x CO (thermodilution)-700 ml. The Doppler method underestimated cardiac output relative to the thermodilution method.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验