Respiratory Medicine Department, School of Medicine, University of Ioannina, Ioannina, Greece.
Respir Res. 2020 Apr 6;21(1):79. doi: 10.1186/s12931-020-01341-z.
Chronic respiratory diseases constitute a considerable part in the practice of pulmonologists and primary care physicians; spirometry is integral for the diagnosis and monitoring of these diseases, yet remains underutilized. The Air Next spirometer (NuvoAir, Sweden) is a novel ultra-portable device that performs spirometric measurements connected to a smartphone or tablet via Bluetooth®.
The objective of this study was to assess the accuracy and validity of these measurements by comparing them with the ones obtained with a conventional desktop spirometer. Two hundred subjects were enrolled in the study with various spirometric patterns (50 patients with asthma, 50 with chronic obstructive pulmonary disease and 50 with interstitial lung disease) as well as 50 healthy individuals.
For the key spirometric parameters in the interpretation of spirometry, i.e. FEV, FVC, FEV/FVC and FEF, Pearson correlation and Interclass Correlation Coefficient were greater than 0.94, exhibiting perfect concordance between the two spirometers. Similar results were observed in an exploratory analysis of the subgroups of patients. Using Bland-Altman plots we have shown good reproducibility in the measurements between the two devices, with small mean differences for the evaluated spirometric parameters and the majority of measurements being well within the limits of agreement.
Our results support the use of Air Next as a reliable spirometer for the screening and diagnosis of various spirometric patterns in clinical practice.
慢性呼吸系统疾病在肺病专家和初级保健医生的实践中占相当大的比例;肺功能测定对于这些疾病的诊断和监测至关重要,但仍未得到充分利用。Air Next 肺功能仪(NuvoAir,瑞典)是一种新型超便携设备,可通过蓝牙®与智能手机或平板电脑连接进行肺功能测量。
本研究的目的是通过将其与传统台式肺功能仪获得的测量值进行比较,评估这些测量值的准确性和有效性。该研究共纳入了 200 名具有不同肺功能模式的受试者(50 名哮喘患者、50 名慢性阻塞性肺疾病患者和 50 名间质性肺疾病患者)以及 50 名健康个体。
对于肺功能测定解释的关键肺功能参数,即 FEV、FVC、FEV/FVC 和 FEF,Pearson 相关系数和组内相关系数均大于 0.94,两种肺功能仪之间具有极好的一致性。在对患者亚组的探索性分析中也观察到了类似的结果。通过 Bland-Altman 图,我们表明两种设备之间的测量具有良好的可重复性,评估的肺功能参数的平均差异较小,并且大多数测量值均在可接受范围内。
我们的结果支持将 Air Next 用作临床实践中各种肺功能模式筛查和诊断的可靠肺功能仪。