Comparison of Goldmann applanation and Ocular Response Analyser tonometry: intraocular pressure agreement and patient preference.

机构信息

Centre for Public Health, Queen's University Belfast, Belfast, UK.

出版信息

Eye (Lond). 2020 Mar;34(3):584-590. doi: 10.1038/s41433-019-0556-2. Epub 2019 Aug 27.

Abstract

OBJECTIVES

To evaluate the agreement between Goldmann applanation tonometry (GAT) and Ocular Response Analyser (ORA) intraocular pressure (IOP) measurements, and patients' preferences.

METHODS

Both eyes of participants in the 'Glaucoma within the Northern Ireland Cohort for the Longitudinal Study of Ageing' (GwNICOLA) were included. Participants underwent GAT by a glaucoma expert and ORA tonometry in a random order. Investigators were masked to measurements between devices. Participants were asked which tonometer, if any, they would prefer. We estimated the 95% limits of agreement (95% LoA) and the variables that influence agreement between tonometers.

RESULTS

There were 228 eyes of 120 participants included in this study. Mean age of participants was 68.0 years (SD 8.79) and 52.5% were female. For GAT-ORA IOPcc the mean difference with GAT (95% CI) was -0.23 mmHg (-0.57 mmHg, 0.11 mmHg) and the 95% LoA (95% CIs) were from 4.82 mmHg (5.15 mmHg, 4.48 mmHg) to -5.28 mmHg (-5.61 mmHg, -4.94 mmHg). 40.8% of eyes had an IOP difference of 2 mmHg or more between GAT and ORA IOPcc. Corneal resistance factor (CRF) as estimated by ORA influenced the agreement between GAT and ORA IOPcc. There were no differences in preference for method of tonometry.

CONCLUSIONS

Although ORA IOPcc measurements with ORA did not show significant bias compared with GAT, the relatively large proportion of measurement differences between ORA IOPcc and GAT that were >2 mmHg indicates that GAT and ORA IOP measurements may not be interchangeable. There were no differences in preference for method of tonometry.

摘要

目的

评估 Goldmann 压平眼压计(GAT)和眼反应分析仪(ORA)眼压(IOP)测量值之间的一致性以及患者的偏好。

方法

纳入“北爱尔兰纵向老龄化研究中的青光眼(GwNICOLA)”参与者的双眼。参与者按照随机顺序接受青光眼专家进行的 GAT 和 ORA 眼压测量。研究人员对两种设备之间的测量值进行了屏蔽。参与者被问到如果有任何选择,他们会更喜欢哪种眼压计。我们估计了两种眼压计之间的 95%一致性界限(95% LoA)和影响一致性的变量。

结果

本研究共纳入 120 名参与者的 228 只眼。参与者的平均年龄为 68.0 岁(SD 8.79),其中 52.5%为女性。对于 GAT-ORA IOPcc,GAT 的平均差值(95% CI)为-0.23mmHg(-0.57mmHg,0.11mmHg),95% LoA(95% CIs)为 4.82mmHg(5.15mmHg,4.48mmHg)至-5.28mmHg(-5.61mmHg,-4.94mmHg)。40.8%的眼睛在 GAT 和 ORA IOPcc 之间的 IOP 差值为 2mmHg 或更高。ORA 估计的角膜阻力因子(CRF)会影响 GAT 和 ORA IOPcc 之间的一致性。两种眼压计的测量方法之间没有偏好差异。

结论

尽管与 GAT 相比,ORA IOPcc 测量值没有显示出明显的偏差,但 ORA IOPcc 和 GAT 之间的测量差异中,有较大比例(>2mmHg)表明 GAT 和 ORA IOP 测量值可能不能互换。两种眼压计的测量方法之间没有偏好差异。