Department of Surgery, Delta Hospital, CHIREC Cancer Institute, Triomflaan 201, 1160, Brussels, Belgium; Department of Radiology, Gasthuisberg University Hospital, Herestraat 49, 3000, Leuven, Belgium.
Alegrete Technology Center, Federal University of Pampa, Av. Tiarajú 810, Ipirabuitã, CEP 97546-550, Alegrete, Rio Grande do Sul, Brazil.
Surg Oncol. 2020 Jun;33:145-157. doi: 10.1016/j.suronc.2020.02.005. Epub 2020 Feb 10.
Radiofrequency ablation (RFA) is increasingly being used to treat unresectable liver tumors. Complete ablation of the tumor and a safety margin is necessary to prevent local recurrence. With current electrodes, size and shape of the ablation zone are highly variable leading to unsatisfactory local recurrence rates, especially for tumors >3 cm. In order to improve predictability, we recently developed a system with four simple electrodes with complete ablation in between the electrodes. This rather small but reliable ablation zone is considered as a building block for matrix radiofrequency ablation (MRFA). In the current study we explored the influence of the electric mode (monopolar or bipolar) and the activation mode (consecutive, simultaneous or switching) on the size and geometry of the ablation zone.
The four electrode system was applied in ex vivo bovine liver. The electric and the activation mode were changed one by one, using constant power of 50 W in all experiments. Size and geometry of the ablation zone were measured. Finite element method (FEM) modelling of the experiment was performed.
In ex vivo liver, a complete and predictable coagulation zone of a 3 × 2 × 2 cm block was obtained most efficiently in the bipolar simultaneous mode due to the combination of the higher heating efficacy of the bipolar mode and the lower impedance by the simultaneous activation of four electrodes, as supported by the FEM simulation.
In ex vivo liver, the four electrode system used in a bipolar simultaneous mode offers the best perspectives as building block for MRFA. These results should be confirmed by in vivo experiments.
射频消融(RFA)越来越多地被用于治疗不可切除的肝肿瘤。为了防止局部复发,需要完全消融肿瘤和安全边界。目前的电极,消融区域的大小和形状高度可变,导致局部复发率不理想,尤其是对于 >3cm 的肿瘤。为了提高预测性,我们最近开发了一种具有四个简单电极的系统,电极之间可以完全消融。这个相当小但可靠的消融区域被认为是矩阵射频消融(MRFA)的一个组成部分。在目前的研究中,我们探讨了电模式(单极或双极)和激活模式(连续、同时或切换)对消融区域大小和形状的影响。
在离体牛肝中应用四电极系统。在所有实验中,均使用 50W 恒功率逐个改变电模式和激活模式。测量消融区域的大小和形状。对实验进行了有限元法(FEM)模拟。
在离体肝中,由于双极模式的加热效率较高,同时激活四个电极导致的阻抗较低,在双极同时激活模式下,最有效地获得了 3×2×2cm 块的完整且可预测的凝固区,这得到了 FEM 模拟的支持。
在离体肝中,双极同时激活模式下使用的四电极系统作为 MRFA 的构建块提供了最佳前景。这些结果应通过体内实验得到证实。