Trifocality Achieved Through Polypseudophakia: Optical Quality and Light Loss Compared With a Single Trifocal Intraocular Lens.

出版信息

J Refract Surg. 2020 Sep 1;36(9):570-577. doi: 10.3928/1081597X-20200715-01.

Abstract

PURPOSE

To compare the optical performance of a twointraocular lens (IOL) system with that of a single capsular bag trifocal IOL.

METHODS

The two-IOL configuration of a monofocal RayOne Aspheric (Rayner Intraocular Lenses, Ltd) and a Sulcoflex Trifocal (Rayner Intraocular Lenses, Ltd) lens was compared in vitro with a single-lens option (RayOne Trifocal; Rayner Intraocular Lenses, Ltd). Two samples of each IOL model were studied with an optical metrology device. The optical quality was assessed using the area under the modulation transfer function (MTF). The impact of the supplementary lens mis-alignment on the MTF was tested. The light loss was also measured using a power meter.

RESULTS

The two-IOL system produced three well-defined focal peaks comparable to those of the single lens. The MTF area of the single- and two-IOL configuration was, respectively, 22.5 and 20.7 at far, 16.4 and 15.4 at intermediate, and 14.9 for each configuration at near. A moderate decentration (up to 0.6 mm) had a minimal effect at intermediate and near on the supplementary lens MTF and no impact at far. A 5° tilt did not alter the MTF. The supplementary lens caused a 1.3% decrease in the optical power.

CONCLUSIONS

The optical quality of the two-IOL system matched that of the single trifocal lens. A low-power supplementary IOL demonstrated high tolerance to misalignment and minimal light attenuation. The reversibility of the two-IOL approach may prove advantageous clinically. [J Refract Surg. 2020;36(9):570-577.].

摘要

目的

比较两 枚人工晶状体(IOL)系统与单枚囊袋三焦点 IOL 的光学性能。

方法

在体外将 RayOne Aspheric(Rayner 眼内晶状体有限公司)和 Sulcoflex 三焦点(Rayner 眼内晶状体有限公司)单焦点 RayOne 晶状体的两 IOL 配置与单透镜选择(RayOne 三焦点;Rayner 眼内晶状体有限公司)进行比较。用光学计量设备研究了每个 IOL 模型的两个样本。使用调制传递函数(MTF)下的面积评估光学质量。测试了补充透镜失准对 MTF 的影响。还使用功率计测量了光损耗。

结果

两 IOL 系统产生了三个定义明确的焦点峰值,与单透镜相当。单透镜和双 IOL 配置的 MTF 面积分别为远为 22.5 和 20.7,中间为 16.4 和 15.4,近为 14.9。适度的偏心(高达 0.6 毫米)在中间和近对补充透镜 MTF 的影响很小,而在远距时则没有影响。5°的倾斜不会改变 MTF。补充透镜会导致光功率降低 1.3%。

结论

两 IOL 系统的光学质量与单三焦点透镜相匹配。低功率补充 IOL 对失准具有很高的容忍度,并且光衰减最小。两 IOL 方法的可逆性在临床上可能具有优势。[J Refract Surg. 2020;36(9):570-577.]。