Visual Performance of Two Diffractive Trifocal Intraocular Lenses: A Randomized Trial.

出版信息

J Refract Surg. 2021 Jul;37(7):460-465. doi: 10.3928/1081597X-20210420-01. Epub 2021 Jul 1.

Abstract

PURPOSE

To compare two trifocal intraocular lenses (IOLs), the RayOne Trifocal RAO603F IOL (closed-loop haptic IOL; Rayner Intraocular Lenses Limited) and the AT Lisa tri 839 MP IOL (plate-haptic IOL; Carl Zeiss Meditec AG), concerning optical and capsular bag performance.

METHODS

Patients scheduled for cataract surgery received either a closed-loop haptic IOL or a plate-haptic IOL in the first eye and the other IOL in the second eye. Three months postoperatively, autorefraction and subjective refraction, uncorrected and corrected distance visual acuity at 4 m, 80 cm, and 40 cm, an objective reading test (Salzburg Reading Desk; SRD Vision), a defocus curve, IOL tilt and decentration, a questionnaire about dysphotopsia, and grading of halos with a halometer were performed.

RESULTS

Eighty-eight eyes of 44 patients were included. Visual acuity was comparable between both IOLs. The closed-loop haptic IOL performed better in the defocus curve at -1.50 diopters (D) (0.08 ± 0.10 vs 0.12 ± 0.09 logMAR; < .01). The plate-haptic IOL had better contrast sensitivity without glare under mesopic and photopic conditions in miosis ( = .0018 and .002, respectively) and mydriasis ( = .017 and .003, respectively). Significant differences were found for less overall subjective disturbance ( = .047) and starbursts ( = .039) for the plate-haptic IOL, but not for the other positive dysphotopsia symptoms.

CONCLUSIONS

Both trifocal IOLs delivered good and comparable visual function with low degrees of disturbing dysphotopsia. The closed-loop haptic IOL was slightly superior in the defocus curve, whereas the plate-haptic IOL was slightly superior concerning contrast sensitivity and positive dysphotopsia. .

摘要

目的

比较两种三焦点人工晶状体(IOL),即 RayOne 三焦点 RAO603F IOL(闭环襻型 IOL;Rayner 眼内晶状体有限公司)和 AT Lisa tri 839 MP IOL(板襻型 IOL;卡尔蔡司 Meditec AG),评估其在光学和囊袋方面的性能。

方法

计划接受白内障手术的患者在第一只眼中接受闭环襻型 IOL 或板襻型 IOL,在第二只眼中接受另一种 IOL。术后 3 个月,进行自动验光和主观验光、4 m、80 cm 和 40 cm 处的未矫正和矫正远视力、客观阅读测试(萨尔茨堡阅读台;SRD Vision)、离焦曲线、IOL 倾斜和偏心、关于视觉干扰的问卷以及使用光晕仪分级光晕。

结果

纳入 44 名患者的 88 只眼。两种 IOL 的视力相当。在-1.50 屈光度(D)(0.08 ± 0.10 与 0.12 ± 0.09 logMAR;<0.01)的离焦曲线上,闭环襻型 IOL 表现更好。在微瞳孔和明瞳孔条件下,板襻型 IOL 的对比敏感度在无眩光时更好(分别为<0.0018 和<0.002,以及分别为<0.017 和<0.003)。在整体主观干扰(=0.047)和星爆(=0.039)方面,板襻型 IOL 的差异有统计学意义,但其他阳性视觉干扰症状则无统计学差异。

结论

两种三焦点 IOL 均提供了良好的、可比的视觉功能,且视觉干扰的程度较低。闭环襻型 IOL 在离焦曲线上略有优势,而板襻型 IOL 在对比敏感度和阳性视觉干扰方面略有优势。