Suppr超能文献

青光眼患者中戈德曼视野计与汉弗莱自动视野计的比较。

A comparison of Goldmann and Humphrey automated perimetry in patients with glaucoma.

作者信息

Trope G E, Britton R

机构信息

Department of Ophthalmology, University of Toronto, Toronto General Hospital, Ontario, Canada.

出版信息

Br J Ophthalmol. 1987 Jul;71(7):489-93. doi: 10.1136/bjo.71.7.489.

Abstract

Humphrey automated threshold perimetry (Program 30-2) was performed on 42 eyes of 25 patients with glaucoma to determine both the sensitivity and specificity of automated perimetry in detecting glaucomatous visual field defects. Automated perimetry sensitivity was 90.38%, while automated perimetry specificity was 91%. Fifty-two patients and a technician took part in a survey to determine their preference for either test. Patients generally preferred having Goldmann perimetry. The technician favoured Humphrey automated perimetry. Program 30-2 on the automated perimeter took 25% longer to perform than Goldmann perimetry.

摘要

对25例青光眼患者的42只眼睛进行了汉弗莱自动阈值视野计检查(程序30 - 2),以确定自动视野计在检测青光眼性视野缺损方面的敏感性和特异性。自动视野计的敏感性为90.38%,特异性为91%。52名患者和一名技术人员参与了一项调查,以确定他们对两种检查方法的偏好。患者总体上更喜欢使用戈德曼视野计。技术人员则青睐汉弗莱自动视野计。自动视野计的程序30 - 2比戈德曼视野计的检查时间长25%。