Suppr超能文献

酶联免疫特异性测定法某些参数的评估

Evaluation of some of the parameters of the enzyme-linked immunospecific assay.

作者信息

Bullock S L, Walls K W

出版信息

J Infect Dis. 1977 Oct;136 Suppl:S279-85. doi: 10.1093/infdis/136.supplement_2.s279.

Abstract

In an attempt to standardize the procedure for the enzyme-linked immunospecific assay, several parameters were examined. It was determined that horseradish peroxidase was the enzyme of choice and that several brands of polystyrene tubes and plates could be used. The test was considerably shortened by omitting long incubation steps previously used for eliminating background fixation of conjugate. The reproducibility of the procedure proved to be excellent, but reactivity curves indicated that use of a single dilution is not adequate for quantitative tests. A "standard" procedure was proposed.

摘要

为使酶联免疫特异性测定程序标准化,对几个参数进行了研究。结果确定辣根过氧化物酶是首选酶,并且可以使用几个品牌的聚苯乙烯试管和酶标板。通过省略先前用于消除结合物背景固定的长时间孵育步骤,该测试被大大缩短。该程序的重现性极佳,但反应曲线表明,单次稀释不足以进行定量测试。因此提出了一种“标准”程序。