Aldridge C, Jones P W, Gibson S, Lanham J, Meyer M, Vannest R, Charles R
J Clin Microbiol. 1977 Oct;6(4):406-13. doi: 10.1128/jcm.6.4.406-413.1977.
An automated, computerized system, the AutoMicrobic System, has been developed for the detection, enumeration, and identification of bacteria and yeasts in clinical specimens. The biological basis for the system resides in lyophilized, highly selective and specific media enclosed in wells of a disposable plastic cuvette; introduction of a suitable specimen rehydrates and inoculates the media in the wells. An automated optical system monitors, and the computer interprets, changes in the media, with enumeration and identification results automatically obtained in 13 h. Sixteen different selective media were developed and tested with a variety of seeded (simulated) and clinical specimens. The AutoMicrobic System has been extensively tested with urine specimens, using a urine test kit (Identi-Pak) that contains selective media for Escherichia coli, Proteus species, Pseudomonas aeruginosa, Klebsiella-Enterobacter species, Serratia species, Citrobacter freundii, group D enterococci, Staphylococcus aureus, and yeasts (Candida species and Torulopsis glabrata). The system has been tested with 3,370 seeded urine specimens and 1,486 clinical urines. Agreement with simultaneous conventional (manual) cultures, at levels of 70,000 colony-forming units per ml (or more), was 92% or better for seeded specimens; clinical specimens yielded results of 93% or better for all organisms except P. aeruginosa, where agreement was 86%. System expansion in progress includes antibiotic susceptibility testing and compatibility with most types of clinical specimens.
已开发出一种自动化的计算机系统——自动微生物系统,用于检测、计数和鉴定临床标本中的细菌和酵母菌。该系统的生物学基础在于冻干的、高度选择性和特异性的培养基,这些培养基被封装在一次性塑料比色皿的孔中;引入合适的标本可使培养基复水并接种到孔中。一个自动化光学系统进行监测,计算机对培养基的变化进行解读,13小时内可自动获得计数和鉴定结果。已开发出16种不同的选择性培养基,并用各种接种(模拟)标本和临床标本进行了测试。自动微生物系统已使用尿检测试剂盒(鉴定包)对尿标本进行了广泛测试,该试剂盒包含用于检测大肠杆菌、变形杆菌属、铜绿假单胞菌、克雷伯菌 - 肠杆菌属、沙雷菌属、弗氏柠檬酸杆菌、D群肠球菌、金黄色葡萄球菌和酵母菌(念珠菌属和光滑球拟酵母菌)的选择性培养基。该系统已对3370份接种尿标本和1486份临床尿标本进行了测试。对于接种标本,每毫升70000个菌落形成单位(或更多)水平下,与同时进行的传统(手工)培养的一致性为92%或更高;临床标本中,除铜绿假单胞菌一致性为86%外,所有生物体的结果为93%或更高。正在进行的系统扩展包括抗生素敏感性测试以及与大多数类型临床标本的兼容性。