Abstract

Bone mineral density (BMD) measurement using dual-energy X-ray absorptiometry (DXA) is considered a diagnostic parameter for osteoporosis by the World Health Organization (WHO). DXA densitometers have different scanning modes for BMD measurements, although the specific scanning modes vary based upon the manufacturer. For DXA machines manufactured by Hologic, which are used globally, a range of scanning modes exist, including but not limited to (in order of decreasing spatial resolution) Array, Fast Array, and Express Array. Only a handful of prior studies have compared the reproducibility of BMD measurements across scan modes. The present study aimed to add to this body of literature by investigating the differences in BMD measured between 3 scanning modes in Hologic DXA machines at 19 different health centers. As part of cross-calibration activities for two multi-center studies in China measuring BMD, the European spine phantom (ESP, 1.000 g/cm) was scanned on 19 different Hologic DXA machines. To measure differences in BMD between the 3 scan modes most commonly found on the Hologic models available (i.e., Array, Fast Array, Express Array), the ESP measurement was performed 10 times for each scan mode on each Hologic DXA machine. One-sample t test was used to compare the average difference between the measured ESP results of the 3 scanning modes at each hospital and reference ESP values. Single factor analysis of variance was performed to compare the average differences between the pairs of scanning modes using the reference ESP. Statistically significant differences between the measured ESP results with reference ESP values were found with each scanning mode at 19 hospitals (all p values <0.05). Consistent with this finding, differences in average BMD between the Array mode and Fast Array mode were invariably the smallest compared to differences seen between the other two pairs of scan modes. Significant differences were observed between average ESP BMD for the Array and Express Array scan modes (0.971 ± 0.013 vs 0.935 ± 0.027, p < 0.001), and between Fast Array and Express Array scan modes (0.972 ± 0.012 vs 0.935 ± 0.027, p < 0.001). However, no significant difference in average ESP BMD was observed between the Array and Fast Array scan modes (0.971 ± 0.013 vs 0.972 ± 0.012, p = 0.997). The selection of ideal scanning mode requires a balance of scanning time, radiation exposure, and measurement accuracy. In this ex vivo study, the Fast Array scanning mode appeared to be a reasonable choice compared with Array and Express Array for BMD measurements by Hologic DXA. Future in vivo studies can help guide the clinical application of these findings.

摘要

采用双能 X 射线吸收法(DXA)测量骨密度(BMD)被世界卫生组织(WHO)视为骨质疏松症的诊断参数。DXA 密度计具有不同的 BMD 测量扫描模式,尽管具体的扫描模式因制造商而异。对于全球使用的霍洛迪克(Hologic)制造的 DXA 仪器,存在多种扫描模式,包括但不限于(按空间分辨率递减顺序排列):Array、Fast Array 和 Express Array。只有少数先前的研究比较了不同扫描模式下 BMD 测量的可重复性。本研究旨在通过研究 19 个不同健康中心的霍洛迪克(Hologic)DXA 仪器上的 3 种扫描模式之间 BMD 的测量差异来补充这方面的文献。作为在中国进行的两项多中心 BMD 测量的交叉校准活动的一部分,欧洲脊柱体模(ESP,1.000 g/cm)在 19 台不同的霍洛迪克(Hologic)DXA 仪器上进行了扫描。为了测量在霍洛迪克(Hologic)模型上最常见的 3 种扫描模式(即 Array、Fast Array 和 Express Array)之间的 BMD 差异,在每台霍洛迪克(Hologic)DXA 仪器上对 ESP 进行了 10 次每种扫描模式的测量。采用单样本 t 检验比较了各医院测量的 ESP 结果与参考 ESP 值之间的平均差异。使用参考 ESP 进行了单因素方差分析,以比较两种扫描模式之间的平均差异。在 19 家医院的每种扫描模式下,ESP 结果与参考 ESP 值之间均存在统计学显著差异(均 P 值 <0.05)。一致的是,与其他两种扫描模式对之间的差异相比,Array 模式和 Fast Array 模式之间的平均 BMD 差异始终最小。Array 和 Express Array 扫描模式之间的平均 ESP BMD 存在显著差异(0.971 ± 0.013 对 0.935 ± 0.027,P < 0.001),Fast Array 和 Express Array 扫描模式之间的平均 ESP BMD 存在显著差异(0.972 ± 0.012 对 0.935 ± 0.027,P < 0.001)。然而,Array 和 Fast Array 扫描模式之间的平均 ESP BMD 无显著差异(0.971 ± 0.013 对 0.972 ± 0.012,P = 0.997)。选择理想的扫描模式需要平衡扫描时间、辐射暴露和测量精度。在这项离体研究中,与 Array 和 Express Array 相比,Fast Array 扫描模式似乎是霍洛迪克(Hologic)DXA 进行 BMD 测量的合理选择。未来的体内研究可以帮助指导这些发现的临床应用。

相似文献

本文引用的文献

4
A Review of Osteoporosis in the Older Adult: An Update.老年骨质疏松症综述:最新进展
Rheum Dis Clin North Am. 2018 Aug;44(3):437-451. doi: 10.1016/j.rdc.2018.03.004. Epub 2018 Jun 13.
6
8
Clinician's Guide to Prevention and Treatment of Osteoporosis.骨质疏松症防治临床指南
Osteoporos Int. 2014 Oct;25(10):2359-81. doi: 10.1007/s00198-014-2794-2. Epub 2014 Aug 15.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验