Lateral spine dual-energy X-ray absorptiometry bone mineral measurement with fan-beam design: effect of osteophytic calcifications on lateral and anteroposterior spine BMD.

作者信息

Franck H, Munz M, Scherrer M, v Lilienfeld-Toal H

机构信息

Clinic Mayenbad, Waldsee, Germany.

出版信息

Rheumatol Int. 1995;15(4):151-4. doi: 10.1007/BF00301772.

Abstract

Recently, fan beam (FB) designs have been made available by several manufacturers (Aloka, Hologic Lunar and Sophar) to measure lumbar spine bone mineral area density (BMD) in both an anteroposterior (AP) and a lateral projection. The present study was performed to evaluate some characteristics of a new dual energy X-ray absorptiometry (DXA) system for supine lateral scans in normals and to study possible advantages for patients with osteophytic calcifications (OC). The precision errors of in vitro and in vivo measurements were estimated by an anthropomorphic phantom and in healthy volunteers. To study the effect of osteoarthritic changes on AP and lateral DXA measurements, BMC (bone mineral content) and BMD were measured in age-matched women (n = 150) with and without OC. Precision errors for lateral BMD in vitro over 1 and 6 months were 0.58/0.67% (slow/fast scan modes) and 0.67/0.77% (slow/fast scan modes), respectively. The short- and mid-term reproducibility of BMD values were 2% and 3.5%, respectively, using the compare function (3.5% and 7.5%, respectively, without the compare facility). The analysis of women with and without OC (n = 150) demonstrated higher mean values for AP BMD (0.892 +/- 0.145 g/cm2) in patients with OC (n = 75) than in normals (0.836 +/- 0.135 g/cm2, n = 75, difference 6.3%). For lateral scans, BMD differed to a minor degree (3.1%) in patients with OC (0.629 +/- 0.133 g/cm2) compared with normals (0.610 +/- 0.117 g/cm2). Corresponding results were obtained in fast FB mode. Furthermore, we found significant (P < 0.0001) correlations between BMD in lateral and AP scans in patients without OC (r = 0.63) and in patients with OC (r = 0.57). Although the FB design facilitated fast AP and lateral scans, the higher precision errors of lateral scans could limit its application in longitudinal studies. The use of compare function should be recommended. However, BMD of lateral scans was less influenced by OC.

摘要

最近,几家制造商(阿洛卡、豪洛捷、通用电气和索法尔)推出了扇形束(FB)设计,用于在前后位(AP)和侧位投影中测量腰椎骨矿物质面积密度(BMD)。本研究旨在评估一种新型双能X线吸收法(DXA)系统用于正常仰卧位侧位扫描的一些特性,并研究对患有骨赘钙化(OC)患者的潜在优势。通过人体模型和健康志愿者估计体外和体内测量的精度误差。为研究骨关节炎变化对AP和侧位DXA测量的影响,对年龄匹配的有和没有OC的女性(n = 150)测量骨矿物质含量(BMC)和BMD。体外侧位BMD在1个月和6个月内的精度误差分别为0.58/0.67%(慢/快扫描模式)和0.67/0.77%(慢/快扫描模式)。使用比较功能时,BMD值的短期和中期再现性分别为2%和3.5%(不使用比较功能时分别为3.5%和7.5%)。对有和没有OC的女性(n = 150)的分析表明,患有OC的患者(n = 75)的AP BMD平均值(0.892 +/- 0.145 g/cm2)高于正常女性(0.836 +/- 0.135 g/cm2,n = 75,差异6.3%)。对于侧位扫描,患有OC的患者(0.629 +/- 0.133 g/cm2)与正常女性(0.610 +/- 0.117 g/cm2)相比,BMD差异较小(3.1%)。在快速FB模式下获得了相应结果。此外,我们发现没有OC的患者(r = 0.63)和患有OC的患者(r = 0.57)的侧位和AP扫描BMD之间存在显著相关性(P < 0.0001)。尽管FB设计便于快速进行AP和侧位扫描,但侧位扫描较高的精度误差可能会限制其在纵向研究中的应用。建议使用比较功能。然而,侧位扫描的BMD受OC的影响较小。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验