Maintaining Image Quality While Reducing Acoustic Noise and Switched Gradient Field Exposure During Lumbar MRI.

机构信息

Department of Nursing, Umeå University, Umeå, Sweden.

Department of Radiation Sciences, Umeå University, Umeå, Sweden.

出版信息

J Magn Reson Imaging. 2021 Jul;54(1):315-325. doi: 10.1002/jmri.27527. Epub 2021 Feb 9.

Abstract

BACKGROUND

MR-generated acoustic noise can contribute to patient discomfort and potentially be harmful. One way to reduce this noise is by altering the gradient output and/or waveform using software optimization. Such modifications might influence image quality and switched gradient field exposure, and different techniques appear to affect sound pressure levels (SPLs) to various degrees.

PURPOSE

To evaluate SPLs, image quality, switched gradient field exposure, and participants' perceived noise levels during two different acoustic noise reduction (ANR) techniques, Quiet Suite (QS) and Whisper Mode (WM), and to compare them with conventional T2-weighted turbo spin echo (T2W TSE) of the lumbar spine.

DESIGN

Prospective.

SUBJECTS

Forty adults referred for lumbar MRI.

FIELD STRENGTH/SEQUENCE: Conventional T2W TSE, T2W TSE with QS, and T2W TSE with WM were acquired at 1.5 T.

ASSESSMENT

Peak SPL (A-weighted decibels, dBA), perceived noise levels (Borg CR10®-scale), signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), three radiologists' qualitative assessments in image quality on an ordinal scale 1-4, switched gradient field exposure (% general public), and gradient currents were measured. Interobserver reliability was reported as percentage agreement.

STATISTICAL TESTS

Repeated measures ANOVA, Friedman's ANOVA, and Wilcoxon's Signed-Rank Test for acoustic noise measurements and image quality assessments.

RESULTS

Mean peak SPLs were 89.9 dBA, 74.3 dBA, and 78.8 dBA for conventional, QS, and WM, respectively (P < 0.05). Participants perceived QS as the quietest and conventional as the loudest sequence (P < 0.05). No qualitative differences in image quality were seen (P > 0.05), although QS showed significantly improved SNR and CNR (P < 0.05). Switched gradient field exposure was reduced by 66% and 48% for QS and WM, respectively.

DATA CONCLUSION

Without degrading image quality, both QS and WM are viable ANR techniques in lumbar T2W TSE. QS provided the lowest SPL, the lowest gradient field exposure and was perceived as the most silent among the three sequences.

LEVEL OF EVIDENCE

1 TECHNICAL EFFICACY STAGE: 5.

摘要

背景

磁共振(MR)产生的噪声可能会引起患者不适,甚至可能造成伤害。一种降低噪声的方法是通过软件优化来改变梯度输出和/或波形。这种修改可能会影响图像质量和切换梯度场暴露,不同的技术似乎会以不同的程度影响声压级(SPL)。

目的

评估两种不同的声学降噪(ANR)技术,即 Quiet Suite(QS)和 Whisper Mode(WM),在 T2 加权涡轮自旋回波(T2W TSE)腰椎扫描中的 SPL、图像质量、切换梯度场暴露和参与者感知噪声水平,并与常规 T2W TSE 进行比较。

设计

前瞻性。

受试者

40 名因腰椎 MRI 就诊的成年人。

磁场强度/序列:在 1.5T 下采集常规 T2W TSE、QS 联合 T2W TSE 和 WM 联合 T2W TSE。

评估

峰值 SPL(A 加权分贝,dBA)、感知噪声水平(Borg CR10®-scale)、信噪比(SNR)、对比噪声比(CNR)、三位放射科医生对图像质量的定性评估(1-4 级有序量表)、切换梯度场暴露(一般公众的百分比)和梯度电流。报告了观察者间一致性作为百分比一致率。

统计检验

用于声学噪声测量和图像质量评估的重复测量方差分析、Friedman ANOVA 和 Wilcoxon 符号秩检验。

结果

常规、QS 和 WM 的平均峰值 SPL 分别为 89.9dBA、74.3dBA 和 78.8dBA(P<0.05)。参与者认为 QS 序列最安静,常规序列最响亮(P<0.05)。图像质量未见明显差异(P>0.05),但 QS 显示 SNR 和 CNR 显著改善(P<0.05)。QS 和 WM 分别使切换梯度场暴露减少了 66%和 48%。

数据结论

在不降低图像质量的情况下,QS 和 WM 都是腰椎 T2W TSE 中可行的 ANR 技术。QS 提供最低的 SPL、最低的梯度场暴露,并且在三种序列中被认为是最安静的。

证据水平

1 技术功效分期:5。