Suppr超能文献

降噪技术对图像质量的影响:一项多厂商研究。

Effect of acoustic noise reduction technology on image quality: a multivendor study.

机构信息

Department of Radiology, Minoh City Hospital, 5-7-1 Kayano, Mino-shi, Osaka, 562-8562, Japan.

Department of Systems and Control Engineering, Faculty of Engineering, Tokushima Bunri University, 1314-1 Shido, Sanuki-shi, Kagawa, 769-2193, Japan.

出版信息

Radiol Phys Technol. 2023 Jun;16(2):235-243. doi: 10.1007/s12194-023-00712-7. Epub 2023 Mar 25.

Abstract

The purpose of this study was to clarify the appropriate use of a combination of pulse sequences and acoustic noise reduction technology in general-purpose brain magnetic resonance imaging. Five pulse sequences commonly used in brain magnetic resonance imaging examinations-turbo spin-echo T2-weighted imaging, T1-weighted fluid-attenuated inversion recovery, T2-weighted fluid-attenuated inversion recovery, diffusion-weighted imaging, and magnetic resonance angiography-were performed on healthy participants at three vendors where acoustic noise reduction technology was available. The results showed that acoustic noise reduction technology reduced sound pressure levels and altered image quality in all pulse sequences across all vendors' magnetic resonance imaging scanners. Although T2-weighted imaging and T1-weighted fluid-attenuated inversion recovery resulted in little image quality degradation, T2-weighted fluid-attenuated inversion recovery, diffusion-weighted imaging, and magnetic resonance angiography had significant image degradation. Therefore, acoustic noise reduction technology should be used with caution.

摘要

本研究旨在阐明在通用脑部磁共振成像中,联合使用脉冲序列和降噪技术的恰当方法。在具备降噪技术的三家厂商的磁共振成像扫描仪上,对健康参与者分别进行了五种常用于脑部磁共振成像检查的脉冲序列——快速自旋回波 T2 加权成像、T1 加权液体衰减反转恢复、T2 加权液体衰减反转恢复、弥散加权成像和磁共振血管造影——的检测。结果表明,降噪技术降低了所有脉冲序列在所有厂商的磁共振成像扫描仪中的声压级并改变了图像质量。尽管 T2 加权成像和 T1 加权液体衰减反转恢复几乎没有导致图像质量下降,但 T2 加权液体衰减反转恢复、弥散加权成像和磁共振血管造影的图像质量有显著下降。因此,应谨慎使用降噪技术。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验