Suppr超能文献

新型骨水泥固定超高交联全聚乙烯髋臼假体的评估:采用放射学立体定位分析法前瞻性随机研究其磨损和固定特征

Evaluation of a new cemented highly cross-linked all-polyethylene cup: a prospective and randomised study assessing wear and fixation characteristics using radiostereometric analysis.

机构信息

Department of Surgical and Perioperative Sciences, Orthopaedics, Umeå University, Umeå, Sweden.

出版信息

Hip Int. 2022 Nov;32(6):779-786. doi: 10.1177/1120700021989991. Epub 2021 Feb 10.

Abstract

BACKGROUND AND PURPOSE

The aim of this prospective, randomised and controlled study was to evaluate the wear and fixation properties of a new cemented highly cross-linked all-polyethylene (HXLPE) cup in comparison with a conventional cemented ultra-high molecular weight polyethylene (ConvPE) cup using radiostereometric analysis (RSA).

PATIENTS AND METHODS

A total of 58 patients (58 hips) with primary osteoarthritis (OA) were enrolled in a randomised controlled trial to receive either a ConvPE cup (control) or HXLPE cup (intervention) with identical geometry. The subjects were randomised in a 1:1 ratio. The primary endpoint was proximal wear measured as femoral head penetration into the cup, secondary outcomes were 3D-wear and annual proximal wear from 1 to 5 years. Cup fixation was measured as movement of the cup in relation to the acetabular bone with proximal migration being the primary outcome measure, 3D-migration and change in inclination as secondary outcomes. The patients were followed for 5 years with RSA performed postoperatively, at 3, 12, 24, and 60 months.

RESULTS

The HXLPE displayed a lower median proximal femoral head penetration compared to ConvPE, with a median difference at 2 years of -0.07 mm (95% CI, -0.10 to -0.04 mm), and -0.19 mm (95% CI, -0.27 to -0.15 mm) at 5 years. Annual proximal wear between 1 and 5 years was 0.03 mm/year for HXLPE and 0.06 mm/year for ConvPE (mean difference 0.05 mm, [95% CI, 0.03-0.07 mm]). Proximal migration, 3D migration and change in inclination was numerically slightly higher for HXLPE, albeit not statistically significant.

CONCLUSIONS

Compared to ConvPE, the HXLPE cup displayed significantly lower polyethylene wear. Cup migration was not statistically significant different.

CLINICALTRIALS.GOV IDENTIFIER: NCT04322799.

摘要

背景与目的

本前瞻性、随机对照研究旨在通过放射学体层摄影术(RSA)评估新型骨水泥交联超高相对分子质量聚乙烯(ConvPE)与传统骨水泥交联全聚乙烯(HXLPE)髋臼杯的磨损和固定性能。

患者与方法

58 例(58 髋)原发性骨关节炎患者入组一项随机对照试验,随机分为接受传统骨水泥交联超高分子量聚乙烯(ConvPE)髋臼杯(对照组)或 HXLPE 髋臼杯(实验组)的两组,两组患者的髋臼杯几何形状相同。以 1:1 的比例随机分组。主要终点为股骨头上杯穿透量,即股骨头近端磨损,次要终点为 1 至 5 年的三维(3D)磨损和每年近端磨损。髋臼杯固定以髋臼杯相对于髋臼骨的移动为测量指标,其中近端迁移为主要观察指标,3D 迁移和倾斜度变化为次要观察指标。患者接受 RSA 随访 5 年,术后、3、12、24 和 60 个月进行随访。

结果

与 ConvPE 相比,HXLPE 组的股骨头上杯穿透量中位数较低,2 年时中位数差值为-0.07mm(95%置信区间,-0.10 至-0.04mm),5 年时中位数差值为-0.19mm(95%置信区间,-0.27 至-0.15mm)。1 至 5 年的每年近端磨损量,HXLPE 组为 0.03mm/年,ConvPE 组为 0.06mm/年(平均差值 0.05mm,95%置信区间,0.03-0.07mm)。尽管髋臼杯近端迁移、3D 迁移和倾斜度变化的数值稍高,但无统计学意义。

结论

与 ConvPE 相比,HXLPE 髋臼杯的聚乙烯磨损明显更低,髋臼杯迁移无统计学差异。

临床试验注册号

NCT04322799。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/3ab1/9726741/47e94e697699/10.1177_1120700021989991-fig1.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验