Rosenfeld Eldar, Rabina Gilad, Barequet Dana, Mimouni Michael, Fischer Naomi, Kurtz Shimon
Department of Ophthalmology, Tel Aviv Sourasky Medical Center, Affiliated to the Sackler Faculty of Medicine, Tel Aviv University, Tel Aviv 64239, Israel.
Department of Ophthalmology, Rambam Health Care Campus, Affiliated to Ruth Rappaport Faculty of Medicine, Technion-Israel Institute of Technology, Haifa 3109601, Israel.
Int J Ophthalmol. 2021 Mar 18;14(3):405-408. doi: 10.18240/ijo.2021.03.12. eCollection 2021.
To evaluate intraocular pressure (IOP) measurements and fluctuations using the iCare ONE rebound tonometer (RT-ONE), during home monitoring, in diagnosed and suspected glaucoma patients.
A retrospective case series of consecutive patients with known glaucoma or glaucoma suspects who were followed-up and treated between January 2016 and January 2017. The study included 80 eyes of 40 patients with a mean age of 59.1±14.6y (range, 24-78). All patients have undergone 4-5d of IOP home monitoring with RT-ONE at morning, noon, afternoon, and night time.
Baseline mean IOP, as measured in the clinic (8 -12 ), was 17.4±5.1 mm Hg, compared to RT-ONE home monitoring mean IOP of 15.6±4.1 mm Hg (=0.002). Mean IOP was significantly lower at noon, afternoon and night times compared to clinic measured IOP and morning measurements (=0.005). IOP peak measured during home monitoring was significantly higher compared to the clinic measured IOP (21.3±5.6 mm Hg and 17.4±5.1 mm Hg, <0.001). IOP peaks during home monitoring demonstrated a majority of 47 peaks during morning measurements, compared to 23 at noon, 19 at afternoon and only 12 at night (<0.001). The home monitoring results led to treatment modification of 44 eyes (55%), treatment regime was insufficient for 40 (50%) eyes.
Home monitoring IOP with RT-ONE can provide good assessment of mean IOP, IOP fluctuations and peaks throughout the hours of the day, which lead to an accurate treatment for glaucoma patients.
使用iCare ONE回弹式眼压计(RT - ONE)评估已确诊和疑似青光眼患者在家中监测期间的眼压(IOP)测量值及波动情况。
对2016年1月至2017年1月期间连续随访并接受治疗的已知青光眼或青光眼疑似患者进行回顾性病例系列研究。该研究纳入了40例患者的80只眼睛,平均年龄为59.1±14.6岁(范围24 - 78岁)。所有患者均使用RT - ONE在家中早晨、中午、下午和夜间进行了4 - 5天的眼压监测。
在诊所测量的基线平均眼压(8 - 12)为17.4±5.1 mmHg,而RT - ONE在家中监测的平均眼压为15.6±4.1 mmHg(P = 0.002)。与诊所测量的眼压和早晨测量值相比,中午、下午和夜间的平均眼压显著降低(P = 0.005)。在家中监测期间测量的眼压峰值明显高于诊所测量的眼压(21.3±5.6 mmHg和17.4±5.1 mmHg,P < 0.001)。在家中监测期间,早晨测量出现了47个眼压峰值,中午为23个,下午为19个,夜间仅为12个(P < 0.001)。家庭监测结果导致44只眼睛(55%)的治疗方案发生改变,40只眼睛(50%)的治疗方案不足。
使用RT - ONE在家中监测眼压可以很好地评估一天中各时段的平均眼压、眼压波动及峰值,从而为青光眼患者提供准确的治疗。