使用 Polar OH1 和 Fitbit Charge 3 可穿戴设备在健康成年人进行轻度、中度、剧烈和冲刺式运动期间测量心率的研究:验证研究。

Measurement of Heart Rate Using the Polar OH1 and Fitbit Charge 3 Wearable Devices in Healthy Adults During Light, Moderate, Vigorous, and Sprint-Based Exercise: Validation Study.

机构信息

Edinburgh Napier University, Edinburgh, United Kingdom.

Division of Biomedical Sciences, University of the Highlands and Islands, Inverness, United Kingdom.

出版信息

JMIR Mhealth Uhealth. 2021 Mar 25;9(3):e25313. doi: 10.2196/25313.

Abstract

BACKGROUND

Accurate, continuous heart rate measurements are important for health assessment, physical activity, and sporting performance, and the integration of heart rate measurements into wearable devices has extended its accessibility. Although the use of photoplethysmography technology is not new, the available data relating to the validity of measurement are limited, and the range of activities being performed is often restricted to one exercise domain and/or limited intensities.

OBJECTIVE

The primary objective of this study was to assess the validity of the Polar OH1 and Fitbit Charge 3 devices for measuring heart rate during rest, light, moderate, vigorous, and sprint-type exercise.

METHODS

A total of 20 healthy adults (9 female; height: mean 1.73 [SD 0.1] m; body mass: mean 71.6 [SD 11.0] kg; and age: mean 40 [SD 10] years) volunteered and provided written informed consent to participate in the study consisting of 2 trials. Trial 1 was split into 3 components: 15-minute sedentary activities, 10-minute cycling on a bicycle ergometer, and incremental exercise test to exhaustion on a motorized treadmill (18-42 minutes). Trial 2 was split into 2 components: 4 × 15-second maximal sprints on a cycle ergometer and 4 × 30- to 50-m sprints on a nonmotorized resistance treadmill. Data from the 3 devices were time-aligned, and the validity of Polar OH1 and Fitbit Charge 3 was assessed against Polar H10 (criterion device). Validity was evaluated using the Bland and Altman analysis, Pearson moment correlation coefficient, and mean absolute percentage error.

RESULTS

Overall, there was a very good correlation between the Polar OH1 and Polar H10 devices (r=0.95), with a mean bias of -1 beats·min and limits of agreement of -20 to 19 beats·min. The Fitbit Charge 3 device underestimated heart rate by 7 beats·min compared with Polar H10, with a limit of agreement of -46 to 33 beats·min and poor correlation (r=0.8). The mean absolute percentage error for both devices was deemed acceptable (<5%). Polar OH1 performed well across each phase of trial 1; however, validity was worse for trial 2 activities. Fitbit Charge 3 performed well only during rest and nonsprint-based treadmill activities.

CONCLUSIONS

Compared with our criterion device, Polar OH1 was accurate at assessing heart rate, but the accuracy of Fitbit Charge 3 was generally poor. Polar OH1 performed worse during trial 2 compared with the activities in trial 1, and the validity of the Fitbit Charge 3 device was particularly poor during our cycling exercises.

摘要

背景

准确、连续的心率测量对于健康评估、体育活动和运动表现非常重要,将心率测量集成到可穿戴设备中扩展了其可及性。尽管光电容积脉搏波技术的使用并不是什么新鲜事,但与测量有效性相关的数据有限,并且进行的活动范围通常限于一个运动领域和/或有限的强度。

目的

本研究的主要目的是评估 Polar OH1 和 Fitbit Charge 3 设备在休息、轻度、中度、剧烈和冲刺型运动期间测量心率的有效性。

方法

共有 20 名健康成年人(9 名女性;身高:平均值 1.73 [标准差 0.1] m;体重:平均值 71.6 [标准差 11.0] kg;年龄:平均值 40 [标准差 10] 岁)自愿参加并书面同意参加研究,研究包括 2 次试验。试验 1 分为 3 个部分:15 分钟的静坐活动、自行车测功机上 10 分钟的自行车运动和电动跑步机上的递增运动至力竭(18-42 分钟)。试验 2 分为 2 个部分:自行车测功机上的 4 次 15 秒最大冲刺和非电动阻力跑步机上的 4 次 30-50 米冲刺。3 个设备的数据进行了时间对齐,并根据 Polar H10(标准设备)评估了 Polar OH1 和 Fitbit Charge 3 的有效性。使用 Bland 和 Altman 分析、Pearson 矩相关系数和平均绝对百分比误差评估有效性。

结果

总体而言,Polar OH1 和 Polar H10 设备之间存在非常好的相关性(r=0.95),平均偏差为-1 次/min,一致性界限为-20 至 19 次/min。与 Polar H10 相比,Fitbit Charge 3 设备低估了心率,一致性界限为-46 至 33 次/min,相关性较差(r=0.8)。两个设备的平均绝对百分比误差均被认为是可接受的(<5%)。Polar OH1 在试验 1 的每个阶段表现良好;然而,试验 2 活动的有效性较差。Fitbit Charge 3 仅在休息和非冲刺跑步机活动中表现良好。

结论

与我们的标准设备相比,Polar OH1 能够准确评估心率,但 Fitbit Charge 3 的准确性通常较差。与试验 1 的活动相比,Polar OH1 在试验 2 期间表现不佳,而 Fitbit Charge 3 设备的有效性在我们的自行车运动中尤其差。