床边即时检测血气分析仪测量脑脊液中乳酸水平的可靠性。

Reliability of a point of care testing blood gas analyzer for measurement of lactate levels in cerebrospinal fluid.

机构信息

Laboratory Medicine Department, Hospital Universitario Santa Lucía, Cartagena, Spain.

出版信息

Scand J Clin Lab Invest. 2024 Feb;84(1):76-78. doi: 10.1080/00365513.2024.2318619. Epub 2024 Feb 20.

Abstract

Analysis of cerebrospinal fluid (CSF), including lactate, is key for diagnosis of acute meningitis. Since blood gas analyzers (BGA) enable rapid and safe blood-lactate measurements, we evaluated the reliability of RAPIDPoint 500 BGA to provide a fast and accurate measure of CSF lactate. In this study, CSF lactate levels were measured by a reference assay and on RAPIDPoint 500 BGA. Comparability was evaluated through difference analysis, using Bland Altman test, and linear regression analysis, using the Passing Bablok test. Agreement rate according to CSF lactate (≥3.5 and <3.5 mmol/L) was calculated using kappa (κ) statistic. Population study included 98 CSF samples. Concerning difference analysis, according to Bland-Altman test, bias was 0.13 mmol/L (CI 95%: -0.26 to 0.52 mmol/L. In regression analysis, according to Passing-Bablok equation a systematic difference between both assays was found. In concordance analysis, the interrate realibility was very high (κ: 0.964). According to our resuls, although a systematic difference was detected when lactate levels were measured on RAPIDPoint 500 BGA, the results from Bland-Altman test and the high agreement rate support that this POCT analyzer could be useful for a early and safe detection of patients with high probability of increased CSF lactate level.

摘要

分析脑脊液(CSF),包括乳酸,是急性脑膜炎诊断的关键。由于血气分析仪(BGA)能够快速、安全地测量血液中的乳酸,因此我们评估了 RAPIDPoint 500 BGA 快速准确测量 CSF 乳酸的可靠性。在这项研究中,我们使用参考方法和 RAPIDPoint 500 BGA 测量 CSF 中的乳酸水平。通过差异分析、Bland-Altman 检验和线性回归分析、Passing-Bablok 检验来评估可比性。根据 CSF 乳酸(≥3.5 和 <3.5mmol/L)计算符合率,使用κ(κ)统计。人群研究包括 98 个 CSF 样本。差异分析显示,根据 Bland-Altman 检验,偏差为 0.13mmol/L(CI95%:-0.26 至 0.52mmol/L)。在回归分析中,根据 Passing-Bablok 方程,发现两种检测方法存在系统差异。一致性分析显示,两种方法的可靠性非常高(κ:0.964)。尽管在使用 RAPIDPoint 500 BGA 测量乳酸水平时发现了系统差异,但 Bland-Altman 检验和高符合率的结果支持这种 POCT 分析仪可用于早期、安全地检测 CSF 乳酸水平升高的可能性较高的患者。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验