Biomed Res Int. 2021 Apr 24;2021:6615560. doi: 10.1155/2021/6615560. eCollection 2021.
The purpose of this study was to evaluate the effect of metal orthodontic brackets on the accuracy of electronic apex locator (EAL). The actual canal lengths (ACL) of 40 mandibular incisor teeth were determined. Then, the teeth were randomly divided into two groups ( = 20). Orthodontic metal brackets were applied in the first group, and no brackets, in the second group. The working length of each tooth was measured with an EAL under 3 test conditions according to the distance between the lip clip and sample tooth. Data were analyzed using one-way repeated measures analysis of variance and Tukey's post hoc tests ( = 0.05). In the bracketed samples, when the lip clip was located at 1 cm and 2 cm from the samples. The mean differences between the EAL measurements and ACLs were statistically higher than those when the samples were located 3 cm from the lip clip ( < 0.05). There were also statistically significant differences between EAL measurements and ACLs in the bracketed samples located 1 and 2 cm from the lip clip ( < 0.05). In the nonbracketed group, the differences between EAL measurements and ACLs were not statistically significant in the samples located 1, 2, and 3 cm from the lip clip ( > 0.05). Use of orthodontic metal brackets can negatively influence the accuracy of the electronic apex locator when the distance between the lip clip and bracket was short. A minimum of 3 cm distance should be kept between the lip clip and tooth in order to make consistent electronic measurements.
本研究旨在评估金属正畸托槽对电子根尖定位仪(EAL)准确性的影响。确定 40 颗下颌切牙的实际根管长度(ACL)。然后,将牙齿随机分为两组(每组 20 颗)。第一组应用正畸金属托槽,第二组不应用托槽。根据唇夹和样本牙之间的距离,使用 EAL 在 3 种测试条件下测量每颗牙的工作长度。使用单向重复测量方差分析和 Tukey 事后检验对数据进行分析( = 0.05)。在带托槽的样本中,当唇夹位于样本 1cm 和 2cm 处时,EAL 测量值与 ACL 之间的平均差异明显高于唇夹位于样本 3cm 处时(<0.05)。当样本位于唇夹 1cm 和 2cm 处时,带托槽样本的 EAL 测量值与 ACL 之间也存在统计学显著差异(<0.05)。在无托槽组中,当唇夹位于样本 1cm、2cm 和 3cm 处时,EAL 测量值与 ACL 之间的差异无统计学意义(>0.05)。当唇夹和托槽之间的距离较短时,正畸金属托槽的使用会对电子根尖定位仪的准确性产生负面影响。为了获得一致的电子测量结果,唇夹和牙齿之间应至少保持 3cm 的距离。