相位稳定的扫频源光学相干断层扫描与主动锁模激光相结合,用于增强视网膜血管造影的对比度。

Phase stable swept-source optical coherence tomography with active mode-locking laser for contrast enhancements of retinal angiography.

机构信息

Advanced Photonics Research Institute, Gwangju Institute of Science and Technology, 123 Cheomdan-gwagiro, Buk-gu, Gwangju, 61005, South Korea.

Harvard Medical School, Boston, MA, 02115, USA.

出版信息

Sci Rep. 2021 Aug 17;11(1):16636. doi: 10.1038/s41598-021-95982-9.

Abstract

Swept-source optical coherence tomography (SS-OCT) is an attractive high-speed imaging technique for retinal angiography. However, conventional swept lasers vary the cavity length of the laser mechanically to tune the output wavelength. This causes sweep-timing jitter and hence low phase stability in OCT angiography. Here, we improve an earlier phase-stabilized, akinetic, SS-OCT angiography (OCTA) method by introducing coherent averaging. We develop an active mode-locking (AML) laser as a high phase-stable akinetic swept source for the OCTA system. The phase stability of the improved system was analyzed, and the effects of coherent averaging were validated using a retina phantom. The effectiveness of the coherent averaging method was further confirmed by comparing coherently and conventionally averaged en face images of human retinal vasculature for their contrast-to-noise ratio, signal-to-noise ratio, and vasculature connectivity. The contrast-to-noise ratio was approximately 1.3 times larger when applying the coherent averaging method in the human retinal experiment. Our coherent averaging method with the high phase-stability AML laser source for OCTA provides a valuable tool for studying healthy and diseased retinas.

摘要

扫频源光学相干断层扫描(SS-OCT)是一种用于视网膜血管成像的有吸引力的高速成像技术。然而,传统的扫频激光器通过机械改变激光腔长来调谐输出波长。这会导致扫频定时抖动,从而导致 OCT 血管成像中的相位稳定性低。在这里,我们通过引入相干平均来改进早期的相位稳定、无运动的 SS-OCT 血管造影(OCTA)方法。我们开发了一种主动锁模(AML)激光作为 OCTA 系统的高相位稳定无运动扫频源。分析了改进系统的相位稳定性,并使用视网膜仿体验证了相干平均的效果。通过比较人视网膜血管的相干和常规平均的正面图像的对比度噪声比、信噪比和血管连通性,进一步证实了相干平均方法的有效性。在人视网膜实验中应用相干平均方法时,对比度噪声比大约增加了 1.3 倍。我们的用于 OCTA 的具有高相位稳定性 AML 激光源的相干平均方法为研究健康和患病的视网膜提供了有价值的工具。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/8586/8371173/5ec4608eebb4/41598_2021_95982_Fig1_HTML.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验