Ophthalmology Department, Faculty of Medicine, Kafkas University, 36100, Merkez, Kars, Turkey.
Ophthalmology Department, Van Training and Research Hospital, Van, Turkey.
Int Ophthalmol. 2022 Mar;42(3):959-968. doi: 10.1007/s10792-021-02079-3. Epub 2021 Oct 22.
To evaluate the performance of precorneal and prelens non-invasive tear break-up time (NIBUT) measurements to determine tear instability in contact lens (CL) wearers.
In this study, 50 eyes of 25 CL wearers were evaluated. Precorneal first and average NIBUTs and fluorescein tear break-up time (FBUT) were measured before wearing CLs in the morning. Those with FBUT less than 10 s were considered to have tear instability. After wearing CL, first and average prelens NIBUTs were measured at the 30th minute. The receiver operating characteristic (ROC) curve was analyzed to evaluate the performance of NIBUT measurements in the diagnosis of tear instability.
The FBUT value had a significant correlation with first and average precorneal NIBUT values (p < 0.001, r = 0.653 and p < 0.001, r = 0.628, respectively). The FBUT value had no correlation with the prelens first and average NIBUT values (p = 0.542 and p = 0.263, respectively). To understand the relationship between the precorneal and prelens NIBUT values measured by the automated method, their correlation was evaluated. There was no significant correlation between the precorneal and prelens NIBUTs (for all; p > 0.05). The area under curve (AUC) in ROC curve for the first and average precorneal NIBUTs were 0.842 (p < 0.001) and 0.810 (p < 0.001), respectively. The AUC values for the first and average prelens NIBUTs at the 30th minute of CL wear were 0.586 and 0.619, respectively (p = 0.317 and p = 0.166, respectively).
Precorneal NIBUT measurements may be useful in diagnosing tear instability. Prelens NIBUT values are not yet capable of adequately defining the tear film dynamics in CL wearers.
评估角膜前和隐形眼镜前非侵入性泪膜破裂时间(NIBUT)测量在确定隐形眼镜(CL)佩戴者泪膜不稳定中的性能。
本研究共评估了 25 名 CL 佩戴者的 50 只眼。在早晨佩戴 CL 前测量角膜前的初始和平均 NIBUT 以及荧光素泪膜破裂时间(FBUT)。FBUT 小于 10 s 的患者被认为存在泪膜不稳定。佩戴 CL 30 分钟后,测量初始和平均隐形眼镜前 NIBUT。通过受试者工作特征(ROC)曲线分析评估 NIBUT 测量在诊断泪膜不稳定中的性能。
FBUT 值与初始和平均角膜前 NIBUT 值有显著相关性(p<0.001,r=0.653 和 p<0.001,r=0.628)。FBUT 值与隐形眼镜前的初始和平均 NIBUT 值无相关性(p=0.542 和 p=0.263)。为了了解自动法测量的角膜前和隐形眼镜前 NIBUT 值之间的关系,评估了它们之间的相关性。角膜前和隐形眼镜前 NIBUT 之间无显著相关性(所有 p 值均>0.05)。ROC 曲线中初始和平均角膜前 NIBUT 的曲线下面积(AUC)分别为 0.842(p<0.001)和 0.810(p<0.001)。CL 佩戴 30 分钟时,初始和平均隐形眼镜前 NIBUT 的 AUC 值分别为 0.586 和 0.619(p=0.317 和 p=0.166)。
角膜前 NIBUT 测量可能有助于诊断泪膜不稳定。隐形眼镜前 NIBUT 值尚不能充分定义 CL 佩戴者的泪膜动力学。