Department of Trauma and Reconstructive Surgery, BG Hospital Bergmannstrost, Halle (Saale), Germany; Department of Trauma, Hand and Reconstructive Surgery, Jena University Hospital - Friedrich Schiller University, Jena, Germany.
Department of Neurosurgery, Jena University Hospital - Friedrich Schiller University, Jena, Germany.
World Neurosurg. 2022 Feb;158:e711-e716. doi: 10.1016/j.wneu.2021.11.043. Epub 2021 Nov 16.
The study sought to investigate the reliability of computed tomography (CT)-derived Hounsfield unit (HU) measurements and ascertain the correlation between HU with quantitative CT (qCT)-derived bone mineral density (BMD) in cases of traumatic thoracolumbar fracture, based on native CT scans.
This study is a retrospective cross-sectional analysis of data sets from patients who received native CT scans and bone mineral density measurements (qCT) of the same vertebral body. Two different CT scanner models were used. The inter-rater reliability of 4 raters, which measured HU in native CT scans, was calculated using intraclass correlation coefficient for absolute agreement (ICC). For the correlation between HU and qCT values, respectively the prediction of qCT based on HU, linear regression was used. Bland-Altman plots were used for visual comparison of predicted and measured qCT values.
In total 305 data sets were analyzed. CT scanner model was found to have no significant impact on HU (P = 0.125). The inter-rater reliability for HU measurements from native CT scans was ICC=0.932 (95% confidence interval 0.919-0.943, P < 0.001). The linear regression showed significant correlation of HU and qCT values for each rater (P < 0.001). The equation for qCT prediction with averaged coefficient and constant is qCT = 0.8 HU + 5. In the Bland-Altman plots no bias of predicted qCT values could be found, but a trend to overestimate predicted higher qCT values and underestimate lower qCT values, respectively.
HU measurement shows very high inter-rater reliability. The HU values correlate closely with qCT BMD values. In summary, it seems that HU measurement is a suitable tool to readily and accurately assess bone quality without further scans or effort in cases of thoracolumbar spinal trauma.
本研究旨在探讨基于原始 CT 扫描的外伤性胸腰椎骨折 CT 衍生的亨氏单位(HU)测量的可靠性,并确定 HU 与定量 CT(qCT)衍生骨密度(BMD)之间的相关性。
本研究是对接受原始 CT 扫描和同一椎体骨密度测量(qCT)的患者数据集进行的回顾性横断面分析。使用了两种不同的 CT 扫描仪模型。使用组内相关系数(ICC)评估 4 名评估者在原始 CT 扫描中测量 HU 的组内一致性。为了分析 HU 与 qCT 值之间的相关性,分别使用线性回归来预测 HU 与 qCT 值之间的关系。Bland-Altman 图用于直观比较预测的和测量的 qCT 值。
共分析了 305 组数据。CT 扫描仪模型对 HU 无显著影响(P=0.125)。原始 CT 扫描 HU 测量的组内一致性 ICC=0.932(95%置信区间 0.919-0.943,P<0.001)。线性回归显示每位评估者的 HU 与 qCT 值均存在显著相关性(P<0.001)。用平均系数和常数预测 qCT 的方程为 qCT=0.8 HU+5。在 Bland-Altman 图中,未发现预测 qCT 值存在偏差,但存在高估预测较高 qCT 值和低估较低 qCT 值的趋势。
HU 测量具有非常高的组内一致性。HU 值与 qCT BMD 值密切相关。总之,HU 测量似乎是一种无需进一步扫描或努力即可快速准确评估胸腰椎脊柱创伤后骨质量的合适工具。