Auffarth Gerd U, Naujokaitis Tadas, Blöck Louise, Daghbashyan Anna, Meis Jan, Augustin Victor A, Khoramnia Ramin, Yildirim Timur M
International Vision Correction Research Centre, Department of Ophthalmology, University of Heidelberg, 69120 Heidelberg, Germany.
Institute of Medical Biometry, University of Heidelberg, 69120 Heidelberg, Germany.
Diagnostics (Basel). 2022 Jan 11;12(1):163. doi: 10.3390/diagnostics12010163.
The aim of this prospective clinical study was to establish and verify an adaptation for axial length (AL) measurement in silicone oil (SO)-filled pseudophakic eyes with a Scheimpflug and partial coherence interferometry (PCI)-based biometer. The AL was measured with a Pentacam AXL (OCULUS Optikgeräte GmbH, Wetzler, Germany) and IOLMaster 700 (Carl Zeiss Meditec, Jena, Germany). The coefficients of variation (CoV) and the mean systematic difference (95% confidence interval (CI)) between the devices were calculated. After implementing a setting for measuring AL in tamponaded eyes with a Pentacam based on data of 29 eyes, another 12 eyes were examined for verification. The mean AL obtained with the Pentacam was 25.53 ± 1.94 mm (range: 21.70 to 30.76 mm), and with IOLMaster, 24.73 ± 1.97 mm (ranged 20.84 to 29.92 mm), resulting in a mean offset of 0.80 ± 0.08 mm (95% CI: 0.77, 0.83 mm), < 0.001. The AL values of both devices showed a strong linear correlation (r = 0.999). Verification data confirmed good agreement, with a statistically and clinically non-significant mean difference of 0.02 ± 0.04 (95% CI: -0.01, 0.05) mm, = 0.134. We implemented a specific adaptation for obtaining reliable AL values in SO-filled eyes with the Pentacam AXL.
这项前瞻性临床研究的目的是使用基于Scheimpflug和部分相干干涉测量法(PCI)的生物测量仪,建立并验证一种适用于硅油(SO)填充的人工晶状体眼轴长(AL)测量的方法。使用Pentacam AXL(德国韦茨勒OCULUS Optikgeräte GmbH公司)和IOLMaster 700(德国耶拿卡尔蔡司医疗技术公司)测量AL。计算了两种设备之间的变异系数(CoV)和平均系统差异(95%置信区间(CI))。基于29只眼睛的数据对Pentacam进行了填充眼AL测量设置后,又对另外12只眼睛进行了验证检查。Pentacam测得的平均AL为25.53±1.94毫米(范围:21.70至30.76毫米),IOLMaster测得的平均AL为24.73±1.97毫米(范围:20.84至29.92毫米),平均偏移为0.80±0.08毫米(95%CI:0.77,0.83毫米),P<0.001。两种设备的AL值显示出很强的线性相关性(r=0.999)。验证数据证实一致性良好,平均差异为0.02±0.04(95%CI:-0.01,0.05)毫米,在统计学和临床上均无显著意义,P=0.134。我们对Pentacam AXL进行了特定调整,以便在SO填充眼中获得可靠的AL值。