硅油填充眼使用光学生物测量法测定眼轴长度的校正因子的开发与验证

Development and Verification of an Adjustment Factor for Determining the Axial Length Using Optical Biometry in Silicone Oil-Filled Eyes.

作者信息

Auffarth Gerd U, Naujokaitis Tadas, Blöck Louise, Daghbashyan Anna, Meis Jan, Augustin Victor A, Khoramnia Ramin, Yildirim Timur M

机构信息

International Vision Correction Research Centre, Department of Ophthalmology, University of Heidelberg, 69120 Heidelberg, Germany.

Institute of Medical Biometry, University of Heidelberg, 69120 Heidelberg, Germany.

出版信息

Diagnostics (Basel). 2022 Jan 11;12(1):163. doi: 10.3390/diagnostics12010163.

Abstract

The aim of this prospective clinical study was to establish and verify an adaptation for axial length (AL) measurement in silicone oil (SO)-filled pseudophakic eyes with a Scheimpflug and partial coherence interferometry (PCI)-based biometer. The AL was measured with a Pentacam AXL (OCULUS Optikgeräte GmbH, Wetzler, Germany) and IOLMaster 700 (Carl Zeiss Meditec, Jena, Germany). The coefficients of variation (CoV) and the mean systematic difference (95% confidence interval (CI)) between the devices were calculated. After implementing a setting for measuring AL in tamponaded eyes with a Pentacam based on data of 29 eyes, another 12 eyes were examined for verification. The mean AL obtained with the Pentacam was 25.53 ± 1.94 mm (range: 21.70 to 30.76 mm), and with IOLMaster, 24.73 ± 1.97 mm (ranged 20.84 to 29.92 mm), resulting in a mean offset of 0.80 ± 0.08 mm (95% CI: 0.77, 0.83 mm), < 0.001. The AL values of both devices showed a strong linear correlation (r = 0.999). Verification data confirmed good agreement, with a statistically and clinically non-significant mean difference of 0.02 ± 0.04 (95% CI: -0.01, 0.05) mm, = 0.134. We implemented a specific adaptation for obtaining reliable AL values in SO-filled eyes with the Pentacam AXL.

摘要

这项前瞻性临床研究的目的是使用基于Scheimpflug和部分相干干涉测量法(PCI)的生物测量仪,建立并验证一种适用于硅油(SO)填充的人工晶状体眼轴长(AL)测量的方法。使用Pentacam AXL(德国韦茨勒OCULUS Optikgeräte GmbH公司)和IOLMaster 700(德国耶拿卡尔蔡司医疗技术公司)测量AL。计算了两种设备之间的变异系数(CoV)和平均系统差异(95%置信区间(CI))。基于29只眼睛的数据对Pentacam进行了填充眼AL测量设置后,又对另外12只眼睛进行了验证检查。Pentacam测得的平均AL为25.53±1.94毫米(范围:21.70至30.76毫米),IOLMaster测得的平均AL为24.73±1.97毫米(范围:20.84至29.92毫米),平均偏移为0.80±0.08毫米(95%CI:0.77,0.83毫米),P<0.001。两种设备的AL值显示出很强的线性相关性(r=0.999)。验证数据证实一致性良好,平均差异为0.02±0.04(95%CI:-0.01,0.05)毫米,在统计学和临床上均无显著意义,P=0.134。我们对Pentacam AXL进行了特定调整,以便在SO填充眼中获得可靠的AL值。