Hirnschall Nino, Varsits Ralph, Doeller Birgit, Findl Oliver
Department of Ophthalmology, VIROS Vienna Institute for Research in Ocular Surgery, A Karl Landsteiner Institute, Hanusch Hospital, Vienna, Austria.
Moorfields Eye Hospital NHS Foundation Trust, London, UK.
Ophthalmol Ther. 2018 Jun;7(1):119-124. doi: 10.1007/s40123-018-0122-1. Epub 2018 Mar 1.
The aim of this study was to find cases in which the axial eye length could not be measured with partial coherence interferometry (PCI) technology and to assess if it could be measured using swept source optical coherence tomography (ss-OCT) technology.
All patients were measured at their pre-assessment visit 1 week prior to cataract surgery using conventional optical biometry (PCI technology, IOLMaster 500, Carl Zeiss Meditec AG, Jena, Germany). Patients in whom one or both eyes could not be measured using PCI technology were invited to participate in the study and to be measured with the ss-OCT (IOL Master 700, Carl Zeiss Meditec AG, Jena, Germany) device.
Altogether, 1226 eyes of 613 patients were measured consecutively, and 78 eyes were not measured successfully with PCI technology. Among those with unsuccessfully measured eyes, 23 patients were willing to participate in the study, and two of those were also unsuccessfully measured with the ss-OCT device (8.7%, 2/23). However, 91.3% (21/23) of the eyes that were unsuccessfully scanned with PCI technology were measurable with the ss-OCT device. The estimated overall rate of unsuccessful scans with the ss-OCT device was 0.5% (6/1226) ([Formula: see text] < 0.01).
ss-OCT technology significantly improves the rate of attainable axial eye length measurements, especially in eyes with posterior subcapsular cataracts, but also in eyes with dense nuclear cataracts, except for white cataracts.
本研究的目的是找出无法使用部分相干干涉测量(PCI)技术测量眼轴长度的病例,并评估是否可以使用扫频源光学相干断层扫描(ss-OCT)技术进行测量。
所有患者在白内障手术前1周的预评估就诊时,使用传统光学生物测量法(PCI技术,IOLMaster 500,德国耶拿卡尔蔡司医疗技术公司)进行测量。无法使用PCI技术测量一只或两只眼睛的患者被邀请参加本研究,并使用ss-OCT(IOL Master 700,德国耶拿卡尔蔡司医疗技术公司)设备进行测量。
总共连续测量了613例患者的1226只眼睛,其中78只眼睛使用PCI技术测量未成功。在这些测量未成功的眼睛中,23例患者愿意参加本研究,其中2例使用ss-OCT设备测量也未成功(8.7%,2/23)。然而,91.3%(21/23)使用PCI技术扫描未成功的眼睛可以使用ss-OCT设备测量。使用ss-OCT设备估计的总体扫描未成功率为0.5%(6/1226)([公式:见正文]<0.01)。
ss-OCT技术显著提高了可获得的眼轴长度测量率,特别是在伴有后囊下白内障的眼睛中,但在伴有致密核性白内障的眼睛中也有提高,白色白内障除外。