Benchmarking the AlignRT surface deformation module for the early detection and quantification of oedema in breast cancer radiotherapy.

作者信息

Sorgato Veronica, Ghazouani Khaoula, Queffelec Yann, Julia Frederic, Clement Sophie, Fric Daniele, Farah Jad

机构信息

Department of Radiotherapy, Daniel Hollard Cancer Institute, Grenoble, France.

School of Medical Physics, Faculty of Medicine, Grenoble Alpes University, France.

出版信息

Tech Innov Patient Support Radiat Oncol. 2022 Jan 15;21:16-22. doi: 10.1016/j.tipsro.2021.12.002. eCollection 2022 Mar.

Abstract

PURPOSE

To determine the accuracy of AlignRT surface deformation module in detecting and quantifying oedema in breast cancer radiotherapy.

MATERIALS AND METHODS

A female torso phantom and water-equivalent boluses of different thicknesses (0.5-1.5 cm) were used. The variation of surface displacement and the percentage of surface within tolerance, as a function of bolus thickness and Region of Interest (ROI) size, were investigated. Additionally, a dynamic phantom was used to study the impact of patient breathing on the swelling estimation. Lastly, a flowchart was derived to alert physicians in the case of breast swelling.

RESULTS

Average displacement value proved to be inversely correlated with ROI size (R > 0.9). As such, for a ROI smaller than the bolus size (2.5x2.5 cm), the average displacement (1.05 cm) provides an accurate estimate of the oedema thickness (within 5%). In opposition, with a clinical ROI, the 1 cm-thick bolus was largely underestimated with an average displacement value of 0.28 cm only. To limit the impact of patient breathing on surface deformation, dynamic surface captures and the use of the corrected patient position should be privileged. Using AlignRT, a clinical workflow for breast swelling follow-up was developed to help in the decision for repeat simulation and dosimetry.

CONCLUSION

The surface deformation module provides an accurate, simple, and radiation-free approach to detect and quantify breast oedema during the course of radiotherapy.

摘要

目的

确定AlignRT表面变形模块在乳腺癌放疗中检测和量化水肿的准确性。

材料与方法

使用女性躯干模型和不同厚度(0.5 - 1.5厘米)的水等效填充物。研究了表面位移的变化以及在公差范围内的表面百分比,作为填充物厚度和感兴趣区域(ROI)大小的函数。此外,使用动态模型研究患者呼吸对肿胀估计的影响。最后,得出一个流程图,以便在乳房肿胀时提醒医生。

结果

平均位移值被证明与ROI大小呈负相关(R > 0.9)。因此,对于小于填充物大小(2.5×2.5厘米)的ROI,平均位移(1.05厘米)可准确估计水肿厚度(误差在5%以内)。相反,对于临床ROI,1厘米厚的填充物被大大低估,平均位移值仅为0.28厘米。为了限制患者呼吸对表面变形的影响,应优先采用动态表面捕捉和使用校正后的患者位置。使用AlignRT,开发了一种用于乳房肿胀随访的临床工作流程,以帮助决定是否重复模拟和剂量测定。

结论

表面变形模块提供了一种准确、简单且无辐射的方法,用于在放疗过程中检测和量化乳房水肿。